beautypg.com

Gige vision timer calculations – ALLIED Vision Technologies Goldeye P-032 SWIR User Manual

Page 48

background image

Appendix

How to install a GigE camera (Bigeye/Pearleye/Goldeye) V2.1.0

48

GigE Vision timer calculations

The figure below shows how an exposure signal is generated.

In order to adjust t

Exp

and t

Dark

of the TREX pulse in each case two parameters

are available:

ExposureTime & ExposureTimeGranularity

DarkTime & DarkTimeGranularity

The granularities are integer values in the range from 0…65535 which define
the minimum and maximum pulse widths. The minimum pulse width is equal to
the time granularity.

Figure 28: GigE Vision timer calculations

t

ExpMIN

μs

 

ExposureTimeGranularity

1

+

 30ns

1000

--------------------------------------------------------------------------------------

=

t

ExpMAX

μs

 

ExposureTimeGranularity

1

+

 30ns

65535

1000

----------------------------------------------------------------------------------------------------------

=

This manual is related to the following products: