3 negative transition filter, 4 event register, 5 event enable register – B&K Precision 4079 - Manual User Manual
Page 83

77
Range: 0 to 131,072. Non integer arguments are rounded before execution.
Command Type: Setting or Query
Setting
Syntax:
:STAT:QUES:PTR
Examples: :STAT:QUES:PTR 2048
Query
Syntax:
:STAT:QUES:PTR?
Response:
NR1
4.13.5.4.3 Negative Transition Filter
:STAT:QUES:NTR
This command is used to set and query the value of the negative transition filter.
Arguments
Type:
NRf
Range:
0 to 131,072. Non integer arguments are rounded before execution
Command Type: Setting or Query
Setting
Syntax:
:STAT:QUES:NTR
Examples: :STAT:QUES:NTR 2048
Query
Syntax:
:STAT:QUES:NTR?
Response: NR1
4.13.5.4.4 Event Register
:STAT:QUES:EVENt?
This query is used to read the event register. Reading the register clears it.
Command Type: Query only
Query
Syntax:
:STATus:QUES:EVEN?
Response: NR1
4.13.5.4.5 Event Enable Register
:STAT:QUES:ENABle
This command is used to set and query the value of the enable register.
Arguments
Type:
NRf
Range:
0 to 131,072. Non integer arguments are rounded before execution.
Command Type: Setting or Query
Setting
Syntax:
:STAT:QUES:ENAB
Examples: :STAT:QUES:ENAB 2048
Query
Syntax:
:STAT:QUES:ENAB?
Response: NR1