beautypg.com

1 pulse period, 2 pulse width, 3 pulse edge – B&K Precision 4079 - Manual User Manual

Page 69: 4 pulse rise

background image

63

4.13.1.12.1 PULSe PERiod

This command sets the pulse period to the specified value.

NOTE: This also sets the pulse frequency.

Arguments

Type:

Numeric

Units:

S, mS, uS, nS

Range: 40nS-2000S

Rounding:

4 digits

Command Type: Setting or Query

Setting

Syntax:

[:SOURce:] PULse: PERiod

[:SOURce:] PULse: PERiod MINimum|MAXimum

Examples:

[:SOURce:] PULse: PERiod 500NS

Query

Syntax:

[:SOURce:] PULse: PERiod?[MINimum|MAXimum]

Response:

NR3

4.13.1.12.2 PULse WIDth

This command pulse width to the specified value.

Arguments

Type:

Numeric

Units:

S, mS, uS, nS

Range:

20 nS minimum; maximum defined by period and transition (see note above)

Rounding:

4 digits

Command Type: Setting or Query

Setting

Syntax:

[:SOURce:]PULse:WIDth< value >

[:SOURce:] PULse:WIDth MINimum|MAXimum

Examples:

[:SOURce:] PULse:WIDth 500NS

Query

Syntax:

[:SOURce:] PULse: WIDth?[MINimum|MAXimum]

Response:

NR3

4.13.1.12.3 PULse EDGe

This command sets both rising and falling edge of the pulse to the specified value.

Arguments

Type:

Numeric

Units:

S, mS, uS, nS

Range:

100 nS minimum; maximum defined by period and width (see note above)

Rounding:

4 digits

Command Type: Setting or Query

Setting

Syntax:

[:SOURce:] PULse:EDGe < value >

[:SOURce:] PULse:EDGe MINimum|MAXimum

Examples:

[:SOURce:] PULse:EDGe 500NS

Query

Syntax:

[:SOURce:] PULse:EDGe?[MINimum|MAXimum]

Response:

NR3

4.13.1.12.4 PULse RISe

This command sets rising edge of the pulse to the specified value.

Arguments

Type:

Numeric

Units:

S, mS, uS, nS

Range:

100 nS minimum; maximum defined by period and width (see note above)

Rounding:

4 digits

This manual is related to the following products: