56
4.13.1.6 Point Rate
:SOURce:PRATe
This command is not used. Point rate setting is available for arbitrary mode only. Please see section 4.13.4.1 for details.
4.13.1.7 AM modulation
The following sections control the AM modulation:
4.13.1.7.1 AM STATe
This command activates or deactivates AM modulation:
Arguments
Type:
Boolean
Command Type: Setting or Query
Setting
Syntax:
[: SOURce:]AM [:STATe]
Examples:
: SOURce:AM :STAT ON
AM OFF
Query
Syntax:
[:SOURce:]AM[:STATe]?
Response:
0|1
4.13.1.7.2 AM DEPTh
This command sets the AM modulation depth in %
Arguments
Type:
Numeric
Units:
none (implied %)
Range:
0 to 100
Rounding:
to integer
Command Type: Setting or Query
Setting
Syntax:
:SOURce:AM:DEPTh
:SOURce:AM:DEPTh
Examples:
AM:DEPTh 50
Query
Syntax:
AM:DEPTh?[
Response:
NR3
4.13.1.7.3 AM SHAPe
This command selects the AM modulating waveform shape
Arguments
Type:
Character
Options:
SINusoid, TRIangle, SQUare
Command Type: Setting or Query
Setting
Syntax:
[:SOURce:]AM:SHAPe
Examples:
[:SOURce:]AM:SHPE SIN
AM:SHAPE TRI
Query
Syntax:
[:SOURce:]AM:SHAPe?
Response:
SIN|TRI|SQU