2 trigger source, 3 burst count – B&K Precision 4079 - Manual User Manual
Page 72

66
This command is used to set the trigger mode. It is not a standard SCPI command.
Arguments
Type:
Character
Options:
CONTinuous
TRIGger
GATE
BURSt
Command Type: Setting or Query
Setting
Syntax:
:TRIGger:MODE
Examples:
:TRIG:MODE CONT
:TRIG:MODE BURS
Query
Syntax:
:TRIGger:MODE?
Response:
CONT|TRIG|GATE|BURS
4.13.3.2 Trigger Source
:TRIGger:SOURce
This command is used to select the trigger source, for use in the Trigger, Gate and Burst trigger modes.
Arguments
Type:
Character
Options:
MANual - Front panel MAN key
BUS - GPIB trigger (GET or *TRG)
INTernal - Internal trigger
EXTernal - External trigger
Command Type: Setting or Query
Setting
Syntax:
:TRIGger:SOURce
Examples:
:TRIG:SOUR BUS
:TRIG:SOUR INT
Query
Syntax: :TRIGger:SOURce?
Response: MAN|BUS|INT|EXT
4.13.3.3 Burst Count
:TRIGger:BURSt
Used to set the number of cycles to be output in the BURST mode. It is not a standard SCPI command.
Arguments
Type:
Numeric
Range:
1 to 999999
Rounding:
to integer value
Command Type: Setting or Query
Setting
Syntax
:TRIGger:BURSt
Examples
:TRIG:BURS 100
:TRIG:BURS MAXIMUM
Query
Syntax:
:TRIGger:BURSt?[
Response:
NR1