8 sync output - marker output - reference output – B&K Precision 4079 - Manual User Manual
Page 102

96
2. Push the MODE key on the unit front panel.
3. Verify that the oscilloscope displays a 1KHz sine wave. Make sure that the Output is ON.
4. Set the external function generator for a square wave output from 0V to 2V at 200 Hz.
5. Connect the function generator output to the unit TRIG IN connector.
6. Set the waveform generator:
Step 1. Set the output mode to triggered
MODE
F2 - TRIG
Step 2. Set the trigger source to internal
F2 - INT
7.
CHECK-Verify that the oscilloscope displays one cycle of sine wave per trigger event.
8.
Set the waveform generator:
Set the output mode to external trigger
F3 - PREV
F3 - EXT
9.
CHECK-Verify that the oscilloscope displays one sine waves per external trigger
10. Set the waveform generator:
Set the output mode to burst.
MODE
F4 - BRST
F2 - INT
11.
CHECK-Verify that the oscilloscope displays two cycles of sine wave per trigger event.
12. If you have a Model 4079 unit, connect the CH2 OUTPUT connector to the oscilloscope, push
CHAN key and repeat steps 2 through 13 for channel 2 (using CH2 control settings and CH2
TRIG IN connector).
5.3.8 Sync Output - Marker Output - Reference Output
Specification: TTL-level pulses at programmed frequency.
Procedure:
2) Connect the SYNC OUT connector to the oscilloscope vertical input.
3) Set the unit to generate sine waves at 1 KHz and
CHECK for a square waveform with
levels of 0V and >4V. on the oscilloscope.
4) Connect the REF OUT connector to the oscilloscope vertical input.
5)
CHECK for a 10 MHz square wave with 0V and >4V levels on the oscilloscope.
6) Connect the MARKER OUT connector to the oscilloscope vertical input.
7) Program the unit to generate an Arbitrary waveform with Marker:
- Press the WAVE key
- Select F5 – ARB
- Select F4 – EDIT
- Select F3 – PREDEF