beautypg.com

Burst wave command – B&K Precision 4052-4055 - Programming Manual User Manual

Page 19

background image

15

EXAMPLE 4

Reads channel two modulate wave parameters that STATE is OFF.

C2:SWWV?

Return:
C2:SWWV STATE,OFF



1.8. Burst Wave Command

DESCRIPTION

Set or get burst wave parameters.


COMMAND SYNTAX

:BursTWaVe

:={C1, C2}

:= {a parameter from the table below.}


Parameters

Value

Description

STATE

Enable or disable burst wave. Note if
you want to set or read burst wave
parameters you must first enable burst
mode.

PRD

When carrier wave is NOISE wave, this
cannot be set. When GATE is selected,
you cannot set this. This can be set
only when trig source is IN (internal).

STPS

When carrier wave is NOISE or PULSE

wave, you can’t set it.

GATE_NCYC

When carrier wave is NOISE, you can’t

set it.

TRSR

When carrier wave is NOISE wave, you
can’t set it. When NCYC was chosen
you can set it.

DLAY

When carrier wave is NOISE wave, you
can’t set it. When NCYC was chosen
you can’t set it.

PLRT

When GATE was chosen you can set it.
When carrier wave is NOISE, it is the
only parameter.

TRMD

When carrier wave is NOISE wave, you
can’t set it. When NCYC was chosen
you can set it. When TRSR is set to
EXT, you can’t set is.