B&K Precision 4052-4055 - Programming Manual User Manual

Page 18

background image

14



where:

:= {ON|OFF}

:={ the same with basic wave frequency}
:={ the same with basic wave frequency}

:= {EXT,INT,MAN}

:= {ON,OFF}

:= {LINE,LOG}

:= {UP,DOWN}

:={ON, OFF}
:={SINE ,SQUARE, RAMP, ARB}
:= { Default unit is "HZ". Minimum value is 1xe-6 HZ,
maximal value depends on the version.}
:={Default unit is "V". Channel one
minimum value 0.004V, Maximal is 6V. Channel two minimum value
0.004V, Maximal is 20V. }
:={ Default unit is "V". maximal value depends on the maximal
value depends on the version.}
:={ 20% to 80%. }
:={ 0% to 100%}


QUERY SYNTAX

:SWeepWaVe?

:={C1, C2}

RESPONSE FORMAT

:={return all parameter of the current sweep wave
parameters.}



EXAMPLE 1

set channel one sweep time to 1 S.

C1:SWWV TIME, 1S


EXAMPLE 2

set channel one sweep stop frequency to 1000hz.

C1: SWWV STOP, 1000HZ



EXAMPLE 3

Reads channel one modulate wave parameters that STATE is ON.

C2:SWWV?

Return:
C2:SWWV
STATE,ON,TIME,1S,STOP,100HZ,START,100HZ,TRSR,MAN,TRMD,OFF,SW
MD,LINE,DIR,UP,CARR,WVTP,SQUARE,FRQ,1000HZ,AMP,4V,OFST,0V,DUTY
,50