Vector network analysis/vector voltmeter, Fsh4/r&s, Fsh-k42/r&s – Atec Rohde-Schwarz-FSH Series User Manual
Page 8: Fsh-k45 option
Version 15.01, July 2013
8
Rohde & Schwarz R&S
®
FSH4/R&S
®
FSH8/R&S
®
FSH13/ R&S
®
FSH20 Spectrum Analyzer
Vector network analysis/vector voltmeter
R&S
®
FSH4/R&S
®
FSH8 model .24/.28 with R&S
®
FSH-K42/R&S
®
FSH-K45 option
Frequency range
R&S
®
FSH4 model .24
300 kHz to 3.6 GHz
R&S
®
FSH8 model .28
300 kHz to 8 GHz
Frequency resolution
1
Hz
Data points
631
Port power
controlled via tracking generator
attenuation
nominal 0 dBm to –40 dBm in 1 dB steps
Reflection measurement
Result formats
measurement mode = vector
magnitude, phase, magnitude + phase,
VSWR, reflection coefficient, Smith chart,
cable loss, group delay, electrical length
measurement mode = vector voltmeter
magnitude + phase, Smith chart
Return loss
Range
selectable
1/2/5/10/20/50/100 dB, linear 100 %
Resolution
0.01
dB
Measurement uncertainty
see figure “Uncertainty of reflection
measurement with the R&S
®
FSH-K42/
R&S
®
FSH-K45 option”
One-port phase
Range
selectable
90/180/360/1000° to 10000° in 1/2/5 steps
Resolution
0.01°
Measurement uncertainty
specifications are based on a matched DUT, RBW = 100 Hz, RF attenuation = 10 dB,
nominal source power = 0 dBm, +20 °C to +30 °C
300 kHz ≤ f ≤ 3.6 GHz
0 dB ≤ return loss < 15 dB
nominal < 3°
15 dB ≤ return loss < 25 dB
nominal < 6°
25 dB ≤ return loss < 35 dB
nominal < 20°
3.6 GHz < f ≤ 8 GHz (R&S
®
FSH8 only)
0 dB ≤ return loss < 15 dB
nominal < 3°
15 dB ≤ return loss < 25 dB
nominal < 6°
25 dB ≤ return loss < 35 dB
nominal < 20°
VSWR
Range
selectable
1 to 1.1, 1.5, 2, 6, 11, 21 or 71
Smith chart
Range
1, zoom Ч 2, Ч 4, Ч 8
Reflection coefficient
mRho
range
1 to 1000 in 1, 2, 5 steps
Corrected directivity
300 kHz ≤ f ≤ 3 GHz
nominal > 43 dB
3 GHz < f ≤ 6 GHz
nominal > 37 dB
6 GHz < f ≤ 8 GHz
nominal > 31 dB
Corrected test port match
300 kHz ≤ f ≤ 3 GHz
nominal > 40 dB
3 GHz < f ≤ 6 GHz
nominal > 37 dB
6 GHz < f ≤ 8 GHz
nominal > 30 dB
Uncertainty of reflection measurement with R&S®FSH-K42/-K45 option
-3
-2
-1
0
1
2
3
0
5
10
15
20
25
30
Return Loss / dB
Mea
sur
em
en
t
un
cer
ta
in
ty
/
d
B
300 kHz ≤ f ≤ 3 GHz
3 GHz < f ≤ 6 GHz
6 GHz < f ≤ 8 GHz
Uncertainty of reflection measurement with the R&S
®
FSH-K42/R&S
®