beautypg.com

Vector network analysis/vector voltmeter, Fsh4/r&s, Fsh-k42/r&s – Atec Rohde-Schwarz-FSH Series User Manual

Page 8: Fsh-k45 option

background image

Version 15.01, July 2013

8

Rohde & Schwarz R&S

®

FSH4/R&S

®

FSH8/R&S

®

FSH13/ R&S

®

FSH20 Spectrum Analyzer

Vector network analysis/vector voltmeter

R&S

®

FSH4/R&S

®

FSH8 model .24/.28 with R&S

®

FSH-K42/R&S

®

FSH-K45 option

Frequency range

R&S

®

FSH4 model .24

300 kHz to 3.6 GHz

R&S

®

FSH8 model .28

300 kHz to 8 GHz

Frequency resolution

1

Hz

Data points

631

Port power

controlled via tracking generator
attenuation

nominal 0 dBm to –40 dBm in 1 dB steps

Reflection measurement
Result formats

measurement mode = vector

magnitude, phase, magnitude + phase,
VSWR, reflection coefficient, Smith chart,
cable loss, group delay, electrical length

measurement mode = vector voltmeter

magnitude + phase, Smith chart

Return loss

Range

selectable

1/2/5/10/20/50/100 dB, linear 100 %

Resolution

0.01

dB

Measurement uncertainty

see figure “Uncertainty of reflection
measurement with the R&S

®

FSH-K42/

R&S

®

FSH-K45 option”

One-port phase

Range

selectable

90/180/360/1000° to 10000° in 1/2/5 steps

Resolution

0.01°

Measurement uncertainty

specifications are based on a matched DUT, RBW = 100 Hz, RF attenuation = 10 dB,
nominal source power = 0 dBm, +20 °C to +30 °C
300 kHz ≤ f ≤ 3.6 GHz

0 dB ≤ return loss < 15 dB

nominal < 3°

15 dB ≤ return loss < 25 dB

nominal < 6°

25 dB ≤ return loss < 35 dB

nominal < 20°

3.6 GHz < f ≤ 8 GHz (R&S

®

FSH8 only)

0 dB ≤ return loss < 15 dB

nominal < 3°

15 dB ≤ return loss < 25 dB

nominal < 6°

25 dB ≤ return loss < 35 dB

nominal < 20°

VSWR

Range

selectable

1 to 1.1, 1.5, 2, 6, 11, 21 or 71

Smith chart

Range

1, zoom Ч 2, Ч 4, Ч 8

Reflection coefficient

mRho

range

1 to 1000 in 1, 2, 5 steps

Corrected directivity

300 kHz ≤ f ≤ 3 GHz

nominal > 43 dB

3 GHz < f ≤ 6 GHz

nominal > 37 dB

6 GHz < f ≤ 8 GHz

nominal > 31 dB

Corrected test port match

300 kHz ≤ f ≤ 3 GHz

nominal > 40 dB

3 GHz < f ≤ 6 GHz

nominal > 37 dB

6 GHz < f ≤ 8 GHz

nominal > 30 dB

Uncertainty of reflection measurement with R&S®FSH-K42/-K45 option

-3

-2

-1

0

1

2

3

0

5

10

15

20

25

30

Return Loss / dB

Mea

sur

em

en

t

un

cer

ta

in

ty

/

d

B

300 kHz ≤ f ≤ 3 GHz

3 GHz < f ≤ 6 GHz

6 GHz < f ≤ 8 GHz

Uncertainty of reflection measurement with the R&S

®

FSH-K42/R&S

®

FSH-K45 option.