beautypg.com

Fsh-k47 1xev-do, Fsh-k47e 1xev-do – Atec Rohde-Schwarz-FSH Series User Manual

Page 16

background image

Version 15.01, July 2013

16

Rohde & Schwarz R&S

®

FSH4/R&S

®

FSH8/R&S

®

FSH13/ R&S

®

FSH20 Spectrum Analyzer

R&S

®

FSH-K47 1xEV-DO

®

BTS pilot channel and EVM measurement

application
R&S

®

FSH-K47E 1xEV-DO

®

BTS PN scanner and time domain power

measurement application

The specifications below apply to the R&S

®

FSH4 and R&S

®

FSH8. They are based on the data sheet specifications of the R&S

®

FSH4

and R&S

®

FSH8, have not been checked separately and are not verified during instrument calibration. Measurement uncertainties are

indicated as 95 % confidence intervals. The specified level measurement errors do not take into account systematic errors due to
reduced signal-to-noise ratio (SNR).

Measurements R&S

®

FSH-K47 R&S

®

FSH-K47E

Spectrum overview

Result summary

RF channel power

Pilot Rho

Carrier frequency error

Traffic activity

Pilot EVM

PN timing (tau)

Peak to average

Pilot power

MAC power

Data power

PN scanner

Detected PN offset

Power per detected PN offset

Burst power

RF channel power

Pilot power


All specifications are valid for RC3, one traffic channel, SNR > 30 dB.

Frequency range

15 MHz to 3.0 GHz

Carrier frequency uncertainty, nominal values
Lock range

±5 kHz

Measurement uncertainty

SNR > 30 dB,
∆f

ref

= uncertainty of reference frequency

< 100 Hz + ∆f

ref

RF channel power
Measurement range

frequency > 15 MHz

preamplifier = off

–60 dBm < P

RF channel

< 20 dBm

preamplifier = on

–75 dBm < P

RF channel

< 20 dBm

Measurement uncertainty

–75 dBm < P

RF channel

< 20 dBm,

ref. level adjusted to RF channel power

< 1 dB, typ. 0.5 dB

Pilot power

SNR > 30 dB

Measurement range

–40 dBm < P

RF channel

< 20 dBm

P

RF channel

– 20 dB < P

PICH

< P

RF channel

Measurement uncertainty

P

RF channel

– 20 dBm < P

CPICH

< P

RF channel

< 1 dB, typ. 0.5 dB

MAC power

SNR > 30 dB

Measurement range

–40 dBm < P

RF channel

< 20 dBm

P

RF channel

– 20 dB < P

SYNC

< P

RF channel

Measurement uncertainty

P

RF channel

– 20 dBm < P

SYNC

< P

RF channel

< 1 dB, typ. 0.5 dB

Data power

SNR > 30 dB

Measurement range

–40 dBm < P

RF channel

< 20 dBm

P

RF channel

– 20 dB < P

SYNC

< P

RF channel

Measurement uncertainty

P

RF channel

– 20 dBm < P

SYNC

< P

RF channel

< 1 dB, typ. 0.5 dB

Pilot EVM

SNR > 30 dB

Measurement range

–40 dBm < P

RF channel

< 20 dBm

1.5 % < EVM < 25 %

Measurement uncertainty

1.5 % < EVM ≤ 10 %

typ. 2.0 %

10 % < EVM < 25 %

typ. 2.5 %

Residual EVM

typ. 2.5 %

Pilot Rho

SNR > 30 dB

Measurement range

–40 dBm < P

RF channel

< 20 dBm

0.9 < Rho < 1

Measurement uncertainty

0.97 < Rho ≤ 1.0

typ. 0.005

0.90 < Rho ≤ 0.97