10

Rohde & Schwarz R&S

®

FSH4/R&S

®

FSH8/R&S

®

FSH13/ R&S

®

FSH20 Spectrum Analyzer

Scalar network analysis

R&S

®

FSH4/R&S

®

FSH8 model .24/.28 without R&S

®

FSH-K42 option

Frequency range

R&S

®

FSH4 model .24

300 kHz to 3.6 GHz

R&S

®

FSH8 model .28

300 kHz to 8 GHz

Frequency resolution

1

Hz

Data points

631

Port power

controlled via tracking generator
attenuation

nominal 0 dBm to –40 dBm in 1 dB steps

Reflection measurement
Result formats

magnitude, VSWR, reflection coefficient

Return loss

range

1/2/5/10/20/50/100 dB, linear 100 %

resolution 0.01

dB

VSWR

range

1 to 2, 6, 11, 21 or 71, selectable

Corrected directivity (20° to 30°)

300 kHz ≤ f ≤ 6 GHz

nominal > 25 dB

6 GHz < f ≤ 8 GHz

nominal > 20 dB

Corrected test port match (20° to 30°)

300 kHz ≤ f ≤ 6 GHz

nominal > 20 dB

6 GHz < f ≤ 8 GHz

nominal > 15 dB

Transmission measurement
Result formats

magnitude

Dynamic range from port 1 to port 2

RF attenuation = 0 dB, tracking generator attenuation = 0 dB, RBW = 1 kHz

300 kHz ≤ f < 6 GHz

> 60 dB, typ. 80 dB

6 GHz ≤ f < 8 GHz

typ. > 40 dB

Dynamic range from port 2 to port 1

RF attenuation = 0 dB, tracking generator attenuation = 0 dB, RBW = 1 kHz

300 kHz ≤ f < 6 GHz

> 70 dB, typ. 90 dB

6 GHz ≤ f < 8 GHz

typ. > 50 dB

Test port match

as specified for tracking generator
output/RF input

0

20

18

16

14

12

10

8

6

4

2

–3

3

2

1

–2

0

–1

Return loss DUT/dB

Measurement uncertainty/dB

Uncertainty of reflection measurement without the R&S

®