Inputs and outputs – Atec Rohde-Schwarz-FSH Series User Manual
Page 7
Version 15.01, July 2013
Rohde & Schwarz R&S
®
FSH4/R&S
®
FSH8/R&S
®
FSH13/ R&S
®
FSH20 Spectrum Analyzer 7
Inputs and outputs
RF input
Impedance
50
Ω
Connector
N
female
VSWR 100
kHz
≤ f ≤ 1 GHz
nominal < 1.5
1 GHz < f ≤ 6 GHz
nominal < 2
6 GHz < f ≤ 20 GHz
nominal < 3
Input attenuator
RF input only
0 dB to 40 dB in 5 dB steps
AF output
AF demodulation types
AM and FM
Connector
3.5 mm mini jack
Output impedance
nominal 32 Ω
Voltage (open circuit)
V
RMS
adjustable from 0 V to > 100 mV
Power sensor
Connector
7-contact female (type Binder 712)
Power sensors supported
see accessories
Tracking generator (models .14/.18/.24/.28 only)
Frequency range
models .14 and .24
100 kHz to 3.6 GHz
models .18 and .28
100 kHz to 8 GHz
Connector
N female, 50 Ω
VSWR 100
kHz
≤ f ≤ 1 GHz
nominal < 1.5
1 GHz < f ≤ 6 GHz
nominal < 2
6 GHz < f ≤ 8 GHz, models .18 and .28
nominal < 3
Output level
tracking generator attenuation = 0 dB
nominal 0 dBm
Tracking generator attenuator
0 dB to 40 dB in 1 dB steps
Dynamic range for isolation
measurements
RF attenuation = 0 dB, tracking generator attenuation = 10 dB, RBW = 1 kHz
100 kHz ≤ f < 300 kHz
> 60 dB, typ. 80 dB
300 kHz ≤ f < 6 GHz
> 70 dB, typ. 90 dB
6 GHz ≤ f < 8 GHz, models .18 and .28 typ. > 50 dB
Reverse power
DC voltage
50 V
CW RF power
+20 dBm (= 0.1 W)
Max. pulse voltage
50 V
Max. pulse energy (10 µs)
1 mWs
External reference, external trigger, DC bias port 2 (BNC 1)
Connector
BNC,
50
Ω
Mode selectable,
R&S
®
FSH4/R&S
®
FSH8, models .24/.28
ext. reference, ext. trigger, DC bias port 2
selectable,
R&S
®
FSH4/R&S
®
FSH8, other models,
R&S
®
FSH13, R&S
®
FSH20
ext. reference, ext. trigger
External reference
required level
0 dBm
frequency 10
MHz
External trigger threshold
low → high transition
2.4 V
high → low transition
0.7 V
DC bias port 2
max. rated input voltage
50 V
max. rated input current
600 mA
IF out, DC bias port 1 (BNC 2)
Connector
BNC,
50
Ω
Mode selectable,
R&S
®
FSH4/R&S
®
FSH8, models .24/.28
IF out, DC bias port 1
R&S
®
FSH4/R&S
®
FSH8, other models,
R&S
®
FSH13, R&S
®
FSH20
IF out
IF out frequency
R&S
®
FSH4/R&S
®
FSH8,
serial number < 105000
21.4 MHz
R&S
®
FSH4/8, serial number ≥ 105000,
R&S
®
FSH13, R&S
®
FSH20
54.4 MHz
DC bias port 1
max. rated input voltage
50 V
max. rated input current
600 mA
AUX
Connector
7-contact female (type Binder 712)