beautypg.com

Inputs and outputs – Atec Rohde-Schwarz-FSH Series User Manual

Page 7

background image

Version 15.01, July 2013

Rohde & Schwarz R&S

®

FSH4/R&S

®

FSH8/R&S

®

FSH13/ R&S

®

FSH20 Spectrum Analyzer 7

Inputs and outputs

RF input
Impedance

50

Connector

N

female

VSWR 100

kHz

≤ f ≤ 1 GHz

nominal < 1.5

1 GHz < f ≤ 6 GHz

nominal < 2

6 GHz < f ≤ 20 GHz

nominal < 3

Input attenuator

RF input only

0 dB to 40 dB in 5 dB steps

AF output
AF demodulation types

AM and FM

Connector

3.5 mm mini jack

Output impedance

nominal 32 Ω

Voltage (open circuit)

V

RMS

adjustable from 0 V to > 100 mV

Power sensor
Connector

7-contact female (type Binder 712)

Power sensors supported

see accessories

Tracking generator (models .14/.18/.24/.28 only)
Frequency range

models .14 and .24

100 kHz to 3.6 GHz

models .18 and .28

100 kHz to 8 GHz

Connector

N female, 50 Ω

VSWR 100

kHz

≤ f ≤ 1 GHz

nominal < 1.5

1 GHz < f ≤ 6 GHz

nominal < 2

6 GHz < f ≤ 8 GHz, models .18 and .28

nominal < 3

Output level

tracking generator attenuation = 0 dB

nominal 0 dBm

Tracking generator attenuator

0 dB to 40 dB in 1 dB steps

Dynamic range for isolation
measurements

RF attenuation = 0 dB, tracking generator attenuation = 10 dB, RBW = 1 kHz

100 kHz ≤ f < 300 kHz

> 60 dB, typ. 80 dB

300 kHz ≤ f < 6 GHz

> 70 dB, typ. 90 dB

6 GHz ≤ f < 8 GHz, models .18 and .28 typ. > 50 dB

Reverse power

DC voltage

50 V

CW RF power

+20 dBm (= 0.1 W)

Max. pulse voltage

50 V

Max. pulse energy (10 µs)

1 mWs

External reference, external trigger, DC bias port 2 (BNC 1)
Connector

BNC,

50

Mode selectable,

R&S

®

FSH4/R&S

®

FSH8, models .24/.28

ext. reference, ext. trigger, DC bias port 2

selectable,
R&S

®

FSH4/R&S

®

FSH8, other models,

R&S

®

FSH13, R&S

®

FSH20

ext. reference, ext. trigger

External reference

required level

0 dBm

frequency 10

MHz

External trigger threshold

low → high transition

2.4 V

high → low transition

0.7 V

DC bias port 2

max. rated input voltage

50 V

max. rated input current

600 mA

IF out, DC bias port 1 (BNC 2)
Connector

BNC,

50

Mode selectable,

R&S

®

FSH4/R&S

®

FSH8, models .24/.28

IF out, DC bias port 1

R&S

®

FSH4/R&S

®

FSH8, other models,

R&S

®

FSH13, R&S

®

FSH20

IF out

IF out frequency

R&S

®

FSH4/R&S

®

FSH8,

serial number < 105000

21.4 MHz

R&S

®

FSH4/8, serial number ≥ 105000,

R&S

®

FSH13, R&S

®

FSH20

54.4 MHz

DC bias port 1

max. rated input voltage

50 V

max. rated input current

600 mA

AUX
Connector

7-contact female (type Binder 712)