Rf characteristics, Frequency, Frequency sweep – Atec Rohde-Schwarz-SMJ100A User Manual
Page 6: Reference frequency
R&S
®
SMJ100A Vector Signal Generator
Version 05.00, December 2007
6
RF characteristics
Frequency
underrange
100 kHz to <300 kHz
R&S
®
SMJ-B103
up to 3 GHz
Range
R&S
®
SMJ-B106
up to 6 GHz
Resolution of setting
0.01 Hz
Resolution of synthesis
standard, fundamental frequency range
750 MHz to 1500 MHz
5 µHz
to within <1 × 10
−7
for f > 200 MHz or
<124 Hz for f < 200 MHz,
with GUI update stopped
after IEC/IEEE bus delimiter
<2 ms, typ. 1.5 ms
in ALC OFF MODE S&H
<4 ms, typ. 2.5 ms
Setting time
1
after trigger pulse in List mode
<450 µs, typ. 300 µs
Phase offset
adjustable in 0.1° steps
Frequency sweep
Operating modes
digital sweep in discrete steps
automatic, step, single,
external single, external step,
manual or external trigger,
linear or logarithmic spacing
Sweep range
full frequency range
linear
full frequency range
Step width
logarithmic
0.01 % to 100 % per step
Reference frequency
Aging
after 30 days of uninterrupted operation
<1 × 10
−9
/day, <1 × 10
−7
/year
Temperature effect
in operating temperature range
±6 × 10
−8
Warm-up time
to nominal thermostat temperature
≤10 min
frequency (approx. sinewave)
10 MHz or external input frequency
level
typ. 5 dBm
Output for internal reference signal
source impedance
50
Ω
frequency
5 MHz, 10 MHz or 13 MHz
maximum deviation
3 × 10
−6
input level, limits
≥−6 dBm, ≤19 dBm
recommended
0 dBm to 19 dBm
Input for external reference
input impedance
50
Ω
sensitivity
typ. 1 × 10
−8
/V to 3 × 10
−8
/V
input voltage
–10 V to +10 V
Electronic tuning from input
AUX I/O
input impedance
10 k
Ω
1
Installation of software that is not authorized by Rohde & Schwarz for use on the R&S
®
SMJ100A or installation of antivirus software can deteriorate the
setting time performance.