General data

Remote control

Systems

IEC/IEEE bus, IEC 60625 (IEEE 488)
Ethernet

Command set

SCPI 1999.5

Connector

IEC/IEEE

24-contact

Amphenol

Ethernet

Western

IEC/IEEE bus address

0 to 30

Interface functions

IEC: SH1, AH1, T6, L4, SR1, RL1, PP1,
DC1, DT1, C0

Operating data

input voltage range, AC, nominal

100 V to 240 V

AC supply frequency

47 Hz to 63 Hz

input current

5.0 A to 1.6 A

Power supply

power factor correction

in line with EN 61000-3-2

in line with EN 55011 class B, EN 61326

EMC

with activated digital I/Q output

in line with EMC directive of EU
(2004/108/EC), applied standard:
EN 61326 (immunity for industrial
environment; class A emissions)

10

Immunity to interfering field strength

up to 10 V/m

operating temperature range

+5 °C to +45 °C
in line with EN 60068-2-1,
EN 60068-2-2

storage temperature range

−20 °C to +60 °C

Environmental conditions

climatic resistance

+40 °C/90 % rel. humidity
in line with EN 60068-2-3

Mechanical resistance
Vibration sinusoidal

5 Hz to 150 Hz,
max. 2 g at 55 Hz,
55 Hz to 150 Hz,
0.5 g const.,
in line with EN 60068-2-6

random

10 Hz to 300 Hz, acceleration 1.2 g (rms),
in line with EN 60068-2-64

Shock

in line with EN 60068-2-27,
MIL-STD-810E
40 g shock spectrum

Electrical safety

in line with EN 61010-1

Dimensions

width × height × depth

435 mm × 192 mm × 460 mm
(17.1 in × 7.6 in × 18.1 in)

Weight

if fully equipped

18 kg (39.7 lb)

Recommended calibration interval

3 years

License information

The firmware of this device contains open source software. Details as well as license agreements can be found in release notes and
operating manual.

10

The instrument complies with the emission requirements stipulated by EN 55011 class A. This means that the instrument is suitable for use in industrial

environments. In line with EN61000-6-4, operation in residential, commercial and business areas or in small-size companies is not covered. Thus, the
instrument may not be operated in residential, commercial and business areas or in small-size companies, unless additional measures are taken to
ensure that EN 610000-6-3 is complied with.