Other options, Ber measurement (r&s, Smj-k80 option) – Atec Rohde-Schwarz-SMJ100A User Manual
Page 31: Bler measurement (r&s
Version 05.00, December 2007
R&S
®
SMJ100A Vector Signal Generator
31
Other options
BER measurement (R&S
®
SMJ-K80 option)
The data supplied by the DUT is compared with a reference pseudo-random bit sequence.
Clock
supplied by DUT; a clock pulse is required
for each valid bit
Clock rate
100 Hz to 60 MHz
PRBS
sequence length
9, 11, 15, 16, 20, 21, 23
pattern ignore
OFF, all 0, all 1
data enable
external
modes
OFF, high, low
restart external
Data
modes ON/OFF
Synchronization time
28 clock cycles
Interface
9-pin D-Sub connector, D-Sub/BNC cable supplied with option
input impedance
1 k
Ω, 50 Ω
trigger threshold
setting range
0 V to 2.50 V
Clock, data, enable, and restart inputs
resolution 0.01
V
Polarity
data, clock, data enable
normal, inverted
Measurement time
selectable by means of maximum number
of data bits or bit errors (max. 2
31
bit
each), continuous measurement
Measurement result
if selected number of data bits or bit errors
is attained
BER in ppm, %, or decade values
Status displays
not synchronized, no clock, no data
BLER measurement (R&S
®
SMJ-K80 option)
In BLER measurement mode, arbitrary data can be provided by the DUT. A signal marking the block’s CRC has to be provided on the
data enable connector of the BER/BLER option.
Clock
supplied by DUT; a clock pulse is required
for each valid bit
Clock rate
100 Hz to 60 MHz
input data
arbitrary
data enable (marking the block’s CRC)
external
Data
modes high,
low
CRC type
CCITT CRC16 (x
16
+ x
12
+ x
5
+ 1)
CRC
CRC bit order
MSB first, LSB first
Synchronization time
1 block
Interface
9-pin D-Sub connector, D-Sub/BNC cable supplied with option
input impedance
1 k
Ω, 50 Ω
trigger threshold
setting range
0 V to 2.50 V
Clock, data, and enable inputs
resolution 0.01
V
Polarity
data, clock, data enable
normal, inverted
Measurement time
selectable by means of maximum number of received blocks or errors (max. 2
31
blocks
each), continuous measurement
Measurement result
if selected number of received blocks or
errors is attained
BLER in ppm, %, or decade values
Status displays
not synchronized, no clock, no data