beautypg.com

Measurement characteristics – Atec Agilent-53200A Series User Manual

Page 6

background image

6

53210A

53220A

53230A

Time interval A to B, B to A

Range

9

N/A

-1 ns to 100,000 s (nom)

-0.5 ns to 100,000 s (min)

Time interval A or B

Range

N/A

2 ns to 100,000 s (min)

Minimum width

N/A

2 ns

Minimum edge repetition
rate

N/A

6 ns

Level & slope

N/A

Auto-level or user selectable

Single-period, pulse-width, rise time, fall time

Range

N/A

0 s to 1000 s

Minimum width

N/A

2 ns

Minimum edge repetition
Rate

N/A

6 ns

Level & slope

N/A

Auto-level or user selectable

Duty

Range

N/A

.000001 to .999999 or 0.0001% to 99.9999%

Minumim width

N/A

2 ns

Level & slope

N/A

Auto-level or user selectable

Phase A to B, B to A

Range

6

N/A

-180.000º to 360.000º

Totalize measurements

Channels

N/A

Ch 1 or Ch 2

Range

9

N/A

0 to 10

15

events

Rate

N/A

0 - 350 MHz

Gating

N/A

Continuous, timed, or external gate input

Gate accuracy is 20 ns

Level measurements

Voltage level - standard
input channels

±5.1 Vpk with 2.5 mV resolution or ±51 Vpk with 25 mV resolution

Microwave power level
(microwave channel option)

0 to 4 relative signal power

Measurement Characteristics

continued