beautypg.com

Timebase – Atec Agilent-53200A Series User Manual

Page 15

background image

15

Timebase

Standard

TCXO

Option 010

Ultra-High Stability OCXO

Aging

1

(spec)

24-hour,

T

CAL

±1 °C

± 0.3 ppb (typ)

30-day, T

CAL

±5 °C

± 0.2 ppm (typ)

± 10 ppb

1-year, T

CAL

±5 °C

± 1 ppm

± 50 ppb

2-year, T

CAL

±5 °C

± 0.5 ppm

± 25 ppb

Temperature (typ)

0 ºC to 55 ºC relative to 25 °C

2

± 1 ppm

± 5 ppb

Factory Calibration uncertainty

Initial factory calibration

3

(typ)

± 0.5 ppm

± 50 ppb

Supplemental characteristics (typ)

5-min. warm-up error

4

± 1 ppm

± 10 ppb

72-hour retrace error

5

< 50 ppb

< 2 ppb

Allan

deviation

t = 1s

1 ppb

0.01 ppb

Timebase Uncertainty = ( Aging + Temperature + Factory Calibration Uncertainty )

1. All Timebase Aging Errors apply only after an initial 30-days of continuous powered operation and for a constant altitude ±100 m.
After the first 1-year of operation, use ½ x (30-day and 1-year) aging rates shown.

2. Additional temperature error corresponding to ‘0

C to 55

C relative to 25

C’ is only included in the time base uncertainty equation if

the temperature of the operating environment is outside the ± 5

C T

CAL

range. If temperature is within T

CAL

± 5

C, that error term

is already included in the aging term above.
3. This term only applies to the factory calibration. Once the instrument is re-calibrated, the term is no longer included.

4. Warm-up error applies when the instrument is powered on in a stable operating environment.

When moved between different operating environments add the Temperature error during the initial 30-minutes of powered operation
5. Retrace error may occur whenever the instrument line-power is removed or whenever the instrument is battery operated and the
battery fully discharges. Retrace error is the residual timebase shift that remains 72-hours after powering-on an instrument that has
experienced a full power-cycle of the timebase. Additional frequency shift errors may occur for instrument exposure to severe

impact shocks >50 g.

Timebase

Front/rear view of 53230A