53210A

53220A

53230A

Math operations

Smoothing (averaging)

1

Selectable 10 (slow), 100 (medium), 1,000 (fast) reading moving average

Selectable filter reset .1% /1000 ppm (fast), .03%/300 ppm (medium), .01%/100 ppm

(slow) change from average

Scaling

mX-b or m(1/X)-b

User settable m and b (offset) values

∆-change

(X-b)/b scaled to %, ppm, or ppb

User settable b (reference) value

Null

(X-b)

User settable b (reference) value

Statistics

1

Mean, standard deviation,

Max, Min, Peak-to-Peak,

count

Mean, standard deviation, Allan deviation

2

, Max, Min, Peak-

to-Peak, count

Limit test

3

Displays PASS/ FAIL message based on user defined Hi/ Lo limit values.

Operation

Individual and simultaneous operation of smoothing, scaling, statistics, and limit test

Graphical display selections

Digits

Numeric result with input level shown

Trend

Strip chart (measurements vs. readings over time)

Selectable screen time

Histogram

Cumulative histogram of measurements; manual reset

HI/LO limit lines shown

Selectable bin and block size

Limit test

Measurement result, tuning bar-graph, and PASS/FAIL message

Markers

Available to read values from trend & histogram displays

Memory

Data log

Guided setup of # of readings/counts;

automatically saves acquisition results to non-volatile memory

Instrument state

Save & recall user-definable instrument setups

Power-off

Automatically saved

Power-on

Selectable power-on to reset (Factory), power-off state or user state

Volatile reading memory

1 M readings (16 MBytes)

Non-volatile internal
memory

75 Mbytes (up to 5 M readings)

USB file system

Front-panel connector for USB memory device

Capability

Store/recall user preferences and instrument states, reading memory,

and bit map displays

Math, Graphing and Memory Characteristics (nom)