Speed characteristics, Meas) – Atec Agilent-53200A Series User Manual
Page 12
12
53210A
53220A
53230A
Maximum measurement speed to internal non-volatile memory
6
: (readings/s)
Timestamp
N/A
N/A
1,000,000
Frequency, period, totalize
300
1000
75,000
Frequency ratio
44,000
Time interval, rise/fall,
width, burst width
N/A
90,000
Duty cycle
N/A
48,000
Phase
N/A
37,000
PRI, PRF
N/A
N/A
75,000
Transfer from memory to PC via:
LAN (sockets)
600,000 readings/sec
LAN (VXI-11)
150,000 readings/sec
USB
800,000 readings/sec
GPIB
22,000 readings/sec
1. These Math operations do not apply for Continuous Totalize or Timestamp measurements.
2. Allan Deviation is only calculated for Frequency and Period measurements. Allan Deviation calculation is available on both 53220A
and 53230A, it is only gap free on 53230A.
3. Limit Test only displays on instrument front panel. No hardware output signal is available.
4. Operating speeds are for a direct connection to a >2.5 GHz dual core CPU running Windows
®
XP Pro SP3 or better with 4 GB RAM and
a 10/100/1000 LAN interface.
5. Throughput data based on gate time. Typical reading throughput assumes ASCII format, Auto level OFF with READ? SCPI command. For
improved reading throughput you should also consider setting (FORM:DATA REAL,64), (DISP OFF), and set fastest gate time available.
6. Maximum 53230A rates represent >= 20 MHz input signals with min gate times, no delays or holdoffs. Measurement rates for the
53210A & 53220A are limited by min gate time. Actual meas rates are limited by the repetition rate of the input being measured.
Speed Characteristics
4
(meas)
continued