Third-order intermodulation distortion (TOI)

Parentheses indicate typical performance

RF (Option 503, 507)

Preamp off

(Two –20 dBm tones at input mixer spaced by

100 kHz, 0 dB attenuation, 20 to 30 °C)

10 to 400 MHz

+10 (+14) dBm

400 MHz to 3 GHz

+13 (+17) dBm

3 to 7.5 GHz

+13 (+15) dBm

MW (Option 513/526)

Preamp off

(Two –20 dBm tones at input mixer spaced by

100 kHz, 0 dB attenuation, 20 to 30 °C)

10 to 500 MHz

+11 dBm, (+15) dBm

500 MHz to 2 GHz

+12 dBm, (+15) dBm

2 to 3 GHz

+11 dBm, (+15) dBm

3 to 7.5 GHz

+12 dBm, (+17) dBm

7.5 to 13.6 GHz

+11 dBm, (+15) dBm

13.6 to 26.5 GHz

+10 dBm, (+14) dBm

Option P03/P07/P13/

P26

Preamp on

(Two –45 dBm tones at the preamp input, spaced by

100 kHz, 0 dB attenuation, 20 to 30 °C)

10 MHz to 26.5 GHz

–8 dBm nominal

Nominal dynamic range at 1 GHz for RF CXA

-130

-120

-110

-100

-90

-80

-70

-60

-80 -70 -60 -50 -40 -30 -20

Mixer level (dBm)

DANL

(1 Hz RBW)
2nd harmonic

distortion
3rd order

intermodulation

DANL and distortion

relative to mixer level (dB)

Nominal dynamic range bands 1-6 for RF CXA

-130

-120

-110

-100

-90

-80

-70

-60

-80 -70 -60 -50 -40 -30 -20

Mixer level (dBm)

DANL at 5.2 GHz

(1 Hz RBW)
2nd harmonic

distortion

at 3.75 GHz
3rd order

intermodulation

at 5.2 GHz

DANL and distortion

relative to mixer level (dB)

Figure 1. Nominal dynamic range for

Options 503 and 507 – Band 0, for second

and third order distortion, 10 MHz to

3 GHz

Figure 2. Nominal dynamic range for

Options 503 and 507 – Bands 1 to 6, for

second and third order distortion, 3 GHz

to 7.5 GHz

Nominal dynamic range for Options 503 and 507