Frequency and time specifications, Frequency range dc coupled ac coupled, Band lo multiple (n) ac coupled – Atec Agilent-N9000A-CXA User Manual
Page 4: Frequency reference
4
Frequency range
DC coupled
AC coupled
Option 503
NA
9 kHz to 3.0 GHz
Option 507
NA
9 kHz to 7.5 GHz
Option 513
9 kHz to 13.6 GHz
10 MHz to 13.6 GHz
Option 526
9 kHz to 26.5 GHz
10 MHz to 26.5 GHz
Band
LO multiple (N)
AC coupled
RF (Option 503, 507)
0
1
9 kHz to 3.0 GHz
1
1
2.95 to 3.80 GHz
2
1
3.70 to 4.55 GHz
3
1
4.45 to 5.30 GHz
4
1
5.20 to 6.05 GHz
5
1
5.95 to 6.80 GHz
6
1
6.70 to 7.50 GHz
Band
LO multiple (N)
AC coupled
MW (Option 513, 526)
0
1
9 kHz to 3.08 GHz
1
2
2.95 to 7.58 GHz
2
2
7.45 to 9.55 GHz
3
2
9.45 to 12.60 GHz
4
2
12.50 to 13.05 GHz
4
4
12.95 to 13.80 GHz
5
4
13.40 to 15.55 GHz
6
4
15.45 to 19.35 GHz
7
4
19.25 to 21.05 GHz
8
4
20.95 to 22.85 GHz
9
4
22.75 to 24.25 GHz
10
4
24.15 to 26.55 GHz
Frequency reference
Accuracy
± [(time since last adjustment x aging rate) + temperature stability + calibration
accuracy]
Aging rate
Option PFR
± 1 x 10
–7
/ year
± 1.5 x 10
–7
/ 2 years
Standard
± 1 x 10
–6
/ year
Temperature stability
20 to 30 °C
Full temperature range
Option PFR
± 1.5 x 10
–8
± 5 x 10
–8
Standard
± 2 x 10
–6
± 2 x 10
–6
Achievable initial calibration accuracy
Option PFR
± 4 x 10
–8
Standard
± 1.4 x 10
–6
Example frequency reference accuracy
(with Option PFR)
1 year after last adjustment
= ± (1 x 1 x 10
–7
+ 5 x 10
–8
+ 4 x 10
–8
)
= ± 1.9 x 10
–7
Residual FM
Option PFR
Standard
≤ 0.25 Hz p-p in 20 ms nominal
≤ 10 Hz p-p in 20 ms nominal
Frequency readout accuracy (start, stop, center, marker)
± (marker frequency x frequency reference accuracy + 0.25 % x span + 5 % x RBW + 2 Hz + 0.5 x horizontal
resolution 1)
1. Horizontal resolution is span/(sweep points – 1).
Frequency and Time Specifications