beautypg.com

Frequency and time specifications, Frequency range dc coupled ac coupled, Band lo multiple (n) ac coupled – Atec Agilent-N9000A-CXA User Manual

Page 4: Frequency reference

background image

4

Frequency range

DC coupled

AC coupled

Option 503

NA

9 kHz to 3.0 GHz

Option 507

NA

9 kHz to 7.5 GHz

Option 513

9 kHz to 13.6 GHz

10 MHz to 13.6 GHz

Option 526

9 kHz to 26.5 GHz

10 MHz to 26.5 GHz

Band

LO multiple (N)

AC coupled

RF (Option 503, 507)

0

1

9 kHz to 3.0 GHz

1

1

2.95 to 3.80 GHz

2

1

3.70 to 4.55 GHz

3

1

4.45 to 5.30 GHz

4

1

5.20 to 6.05 GHz

5

1

5.95 to 6.80 GHz

6

1

6.70 to 7.50 GHz

Band

LO multiple (N)

AC coupled

MW (Option 513, 526)

0

1

9 kHz to 3.08 GHz

1

2

2.95 to 7.58 GHz

2

2

7.45 to 9.55 GHz

3

2

9.45 to 12.60 GHz

4

2

12.50 to 13.05 GHz

4

4

12.95 to 13.80 GHz

5

4

13.40 to 15.55 GHz

6

4

15.45 to 19.35 GHz

7

4

19.25 to 21.05 GHz

8

4

20.95 to 22.85 GHz

9

4

22.75 to 24.25 GHz

10

4

24.15 to 26.55 GHz

Frequency reference

Accuracy

± [(time since last adjustment x aging rate) + temperature stability + calibration

accuracy]

Aging rate

Option PFR

± 1 x 10

–7

/ year

± 1.5 x 10

–7

/ 2 years

Standard

± 1 x 10

–6

/ year

Temperature stability

20 to 30 °C

Full temperature range

Option PFR

± 1.5 x 10

–8

± 5 x 10

–8

Standard

± 2 x 10

–6

± 2 x 10

–6

Achievable initial calibration accuracy

Option PFR

± 4 x 10

–8

Standard

± 1.4 x 10

–6

Example frequency reference accuracy

(with Option PFR)

1 year after last adjustment

= ± (1 x 1 x 10

–7

+ 5 x 10

–8

+ 4 x 10

–8

)

= ± 1.9 x 10

–7

Residual FM

Option PFR

Standard

≤ 0.25 Hz p-p in 20 ms nominal

≤ 10 Hz p-p in 20 ms nominal

Frequency readout accuracy (start, stop, center, marker)

± (marker frequency x frequency reference accuracy + 0.25 % x span + 5 % x RBW + 2 Hz + 0.5 x horizontal

resolution 1)

1. Horizontal resolution is span/(sweep points – 1).

Frequency and Time Specifications