beautypg.com

Atec Agilent-N9000A-CXA User Manual

Page 10

background image

10

Nominal dynamic range at 1 GHz for RF CXA

-130

-120

-110

-100

-90

-80

-70

-60

-80 -70 -60 -50 -40 -30 -20

Mixer level (dBm)

DANL

(1 Hz RBW)
2nd harmonic

distortion
3rd order

intermodulation

DANL and distortion

relative to mixer level (dB)

Figure 3. Nominal dynamic range for

option 513/526, for second and third order

distortion, 100 MHz to 26.5 GHz

Nominal dynamic range for Options 513 and 526

Figure 4. Nominal phase noise at different center frequencies for RF CXA (Option 503 or 507)

Phase noise

1

Offset

Specification

Typical

Noise sidebands (20 to 30 °C, CF = 1 GHz)

RF (Option 503, 507)

1 kHz

10 kHz

100 kHz

1 MHz

10 MHz

–94 dBc/Hz

–99 dBc/Hz

–102 dBc/Hz

–120 dBc/Hz

–98 dBc/Hz nominal

–102 dBc/Hz

–104 dBc/Hz

–121 dBc/Hz

–143 dBc/Hz nominal

1. For nominal phase noise values with the RF CXA (Option 503 or 507), refer to Figure 4.