I/q analyzer – Atec Agilent-N9000A-CXA User Manual
Page 18
18
Frequency
Frequency span
Standard instrument
Option B25
10 Hz to 10 MHz
10 Hz to 25 MHz
Resolution bandwidth (spectrum measurement)
Range
Overall
Span = 1 MHz
Span = 10 kHz
Span = 100 Hz
100 mHz to 3 MHz
50 Hz to 1 MHz
1 Hz to 10 kHz
100 mHz to 100 Hz
Window shapes
Flat top, Uniform, Hanning, Gaussian, Blackman, Blackman-Harris, Kaiser Bessel (K-B 70 dB, K-B 90 dB and K-B 110 dB)
Analysis bandwidth
Standard instrument
Option B25
10 Hz to 10 MHz
10 Hz to 25 MHz
IF frequency response (standard 10 MHz IF path)
IF frequency response (demodulation and FFT response relative to the center frequency, 20 to 30 °C)
Center frequency (GHz)
Span (MHz)
Max. error
RMS (nominal)
≤ 3.0
3.0 < f ≤ 7.5
≤ 10
≤ 10
± 0.45 dB
± 0.45 dB
0.03 dB
0.25 dB
IF phase linearity (deviation from mean phase linearity, nominal)
Center frequency (GHz)
Span (MHz)
Peak-to-peak
RMS
≤ 3.0
3.0 < f ≤ 7.5
≤ 10
≤ 10
± 0.5 °
± 1.5 °
0.2 °
0.4 °
Data acquisition (standard 10 MHz IF path)
Time record length
Sample rate
ADC resolution
4,000,000 IQ sample pairs
30 MSa/s
14 Bits
Option B25 25 MHz analysis bandwidth
IF frequency response (demodulation and FFT response relative to the center frequency, 20 to 30 °C)
Center frequency (GHz)
Span (MHz)
Max. error
RMS (nominal)
≤ 3.0
3.0 < f ≤ 7.5
10 to ≤ 25
10 to ≤ 25
± 0.45 dB
± 0.45 dB
0.03 dB
0.65 dB
IF phase linearity (deviation from mean phase linearity, nominal)
Center frequency (GHz)
Span (MHz)
Peak-to-peak
RMS
0.02 ≤ f < 3.0
3.0 < f ≤ 7.5
10 to ≤ 25
10 to ≤ 25
± 0.8 °
± 1.5 °
± 0.3 °
± 0.4 °
Data acquisition (B25 IF path)
Time record length
IQ analyzer
Sample rate
ADC resolution
4,000,000 IQ sample pairs
90 MSa/s
14 Bits
I/Q Analyzer