Tracking generator specifications, Output frequency, Output power level – Atec Agilent-N9000A-CXA User Manual
Page 13: Output power sweep, Maximum safe reverse level, Phase noise, Spurious outputs (0 dbm output), Dynamic range, Output vswr
13
Output frequency
Frequency range
Option T03
1
Option T06
1
9 kHz to 3 GHz
9 kHz to 6 GHz
Resolution
1 Hz
Output power level
Range
−50 to 0 dBm
Resolution
0.1 dB
Absolute accuracy
(at 50 MHz, −10 dBm, 20 to 30 °C)
± 0.55 dB
Output flatness
(referenced to 50 MHz, −10 dBm, 20 to 30 °C)
Specification
95th percentile (≈ 2σ)
9 kHz to 100 kHz
100 kHz to 3.0 GHz
3.0 GHz to 6.0 GHz
± 1.5 dB
± 1.2 dB
± 1.5 dB
± 1.2 dB
± 0.8 dB
± 1.2 dB
Level accuracy
9 kHz to 100 kHz
100 kHz to 3.0 GHz
3.0 GHz to 6.0 GHz
± 1.0 dB nominal
± 0.5 dB nominal
± 0.8 dB nominal
Output power sweep
Range
−50 to 0 dBm
Resolution
0.1 dB
Maximum safe reverse level
Average total power
+30 dBm (1 W)
AC coupled
± 50 Vdc
Phase noise
Noise sidebands (CF = 1 GHz)
Offset
10 kHz
100 kHz
1 MHz
−102 dBc/Hz nominal
−104 dBc/Hz nominal
−120 dBc/Hz nominal
Spurious outputs (0 dBm output)
Harmonic Spurs
100 kHz to 3 GHz
3 GHz to 6 GHz
< −35 dBc
< −30 dBc
Non-harmonic spurs
9 kHz to 10 MHz
10 MHz to 6 GHz
< −35 dBc
< −35 dBc nominal
Dynamic range
Maximum output power – displayed
average noise level
110 dBc nominal
Output VSWR
9 kHz to 6 GHz
< 1.5:1 nominal
1. Not available on microwave CXA (Option 513 or 526).
Tracking Generator Specifications