Amplitude accuracy and range specifications, Amplitude range, Maximum safe input level – Atec Agilent-N9000A-CXA User Manual
Page 6: Display range
6
Amplitude range
Measurement range
RF (Option 503, 507)
Preamp off
100 kHz to 1 MHz
Displayed average noise level (DANL) to +20 dBm
1 MHz to 7.5 GHz
Displayed average noise level (DANL) to +23 dBm
Preamp on
100 kHz to 7.5 GHz
Displayed average noise level (DANL) to +15 dBm
MW (Option 513/526)
Preamp off
100 kHz to 26.5 GHz
Displayed average noise level (DANL) to +23 dBm
Preamp on
100 kHz to 26.5 GHz
Displayed average noise level (DANL) to +23 dBm
Input attenuator range
RF (Option 503, 507)
Standard
Option FSA
0 to 50 dB in 10 dB steps
0 to 50 dB in 2 dB steps
MW (Option 513, 526)
Standard
Option FSA
0 to 70 dB in 10 dB steps
0 to 70 dB in 2 dB steps
Maximum safe input level
Average total power
RF (Option 503, 507)
+30 dBm (1 W)
Input attenuation ≥ 20 dB, preamp off
10 dBm (10 mW)
Input attenuation ≥ 20 dB, preamp on
MW (Option 513, 526)
+30 dBm (1 W)
Input attenuation ≥ 10 dB, preamp off
+30 dBm (1 W)
Input attenuation ≥ 20 dB, preamp on
Peak pulse power
+50 dBm (100 W)
< 10 μs pulse width, < 1 % duty cycle, input attenuation ≥ 30 dB
DC volts
RF (Option 503, 507)
AC coupled
± 50 Vdc
MW (Option 513, 526)
AC coupled
± 50 Vdc
DC coupled
± 0.2 Vdc
Display range
Log scale
0.1 to 1 dB/division in 0.1 dB steps
1 to 20 dB/division in 1 dB steps (10 display divisions)
Linear scale
10 divisions
Scale units
dBm, dBmV, dBμV, dBmA, dBμA, V, W, A
Frequency response
Specification
95th percentile (≈ 2σ)
(10 dB input attenuation, 20 to 30 °C, σ = nominal standard deviation)
RF (Option 503, 507)
9 kHz to 10 MHz
± 0.60 dB
± 0.45 dB
10 MHz to 3 GHz
± 0.75 dB
± 0.55 dB
3 to 5.25 GHz
± 1.45 dB
± 1.00 dB
5.25 to 7.5 GHz
± 1.65 dB
± 1.20 dB
MW (Option 513, 526)
9 kHz to 10 MHz
± 0.8 dB
± 0.5 dB
10 MHz to 3 GHz
± 0.65 dB
± 0.4 dB
3 to 7.5 GHz
± 1.5 dB
± 0.5 dB
7.5 to 13.6 GHz
± 2.0 dB
± 0.8 dB
13.6 to 19 GHz
± 2.0 dB
± 1.0 dB
19 to 26.5 GHz
± 2.5 dB
± 1.3 dB
Preamp on
RF (Option 503, 507)
(P03, P07)
100 kHz to 3 GHz
± 0.70 dB
3 to 5.25 GHz
± 0.85 dB
5.25 to 7.5 GHz
± 1.35 dB
MW (Option 513, 526)
(P03, P07, P13, P26)
100 kHz to 3 GHz
± 0.7 dB
3 to 13.6 GHz
± 1.0 dB
13.6 to 19 GHz
± 1.1 dB
19 to 26.5 GHz
± 2.5 dB
Amplitude Accuracy and Range Specifications