Temperature range
Operating
5 to 50 °C
Storage
–40 to 70 °C
EMC
Complies with European EMC Directive 2004/108/EC
• IEC/EN 61326-1 or IEC/EN 61326-2-1
• CISPR Pub 11 Group 1, class A
• AS/NZS CISPR 11:2002
• ICES/NMB-001
This ISM device complies with Canadian ICES-001
Cet appareil ISM est conforme à la norme NMB-001 du Canada
Safety
Complies with European Low Voltage Directive 73/23/EEC, amended by 93/68/EEC
• IEC/EN 61010-1 2nd Edition
• Canada: CSA C22.2 No. 61010-1
• USA: UL 61010-1 2nd Edition
Audio noise
Acoustic noise emission
Geraeuschemission
LpA < 70 dB
LpA < 70 dB
Operator position
Am Arbeitsplatz
Normal position
Normaler Betrieb
Per ISO 7779
Nach DIN 45635 t.19
Environmental stress
Samples of this product have been type tested in accordance with the Agilent Environmental Test Manual and verified to be robust
against the environmental stresses of storage, transportation, and end-use; those stresses include, but are not limited to, temperature,
humidity, shock, vibration, altitude, and power line conditions; test methods are aligned with IEC 60068-2 and levels are similar to
MILPRF-28800F Class 3.
Power requirements
Voltage and frequency (nominal)
100 to 120 V, 50/60/400 Hz
220 to 240 V, 50/60 Hz
Power consumption
On
Standby
270 W maximum
20 W
Display
Resolution
1024 x 768, XGA
Size
213 mm (8.4 in.) diagonal (nominal)
Data storage
Internal
80 GB nominal (removable solid state drive)
External
Supports USB 2.0 compatible memory devices
Weight (without options)
Net
15.4 kg (34.0 lbs)
Shipping
27.4 kg (60.4 lbs)
Dimensions
Height
177 mm (7.0 in)
Width
426 mm (16.8 in)
Length
368 mm (14.5 in)
Warranty
The CXA signal analyzer is supplied with a one-year warranty
Calibration cycle
The recommended calibration cycle is one year; calibration services are available through Agilent service centers
General Specifications