beautypg.com

Ber measurement function – Atec Anritsu-MG3710A User Manual

Page 34

background image

34

Data Sheet

l

MG3710A

BER Measurement Function

with MG3710A-021

Connector

Connector: AUX Connector (Rear panel)

Level: TTL

Input Signal

Data, Clock, Enable

Input Bit Rate

100 bps to 40 Mbps

Measurable Pattern

Repeat PN9, PN11, PN15, PN20, PN23, ALL0, ALL1, 01

PN9fix, PN11fix, PN15fix, PN20fix, PN23fix, User Define

Synchronization Establishing Condition

PN Signal: PN order × 2-bit error free

PNfix Signal: Syncs with PN signal at PN order × 2-bit error free;

Syncs with Pnfix signal at PN order error free from Pnfix signal header bit

Repeat ALL0, ALL1, 01: 10-bit Error Free

UserDefine: 8-bit to 1024-bit (variable) error free; can select header bit for Sync detection

Re-synchronization Judgment

x/y

x: Number of error bits in y bit (Setting range: 1 to y/2)

y: Number of measurement bits (select from 500, 5000 and 50000)

Measurable Bit

≤ 2

32

– 1 bit

Measurable Error Bit Count

≤ 2

32

– 1 bit

Measurement End

Number of measurement bits, Number of measurement error bits

Automatic Re-synchronization

Can be toggled on and off

Re-synchronization

Count Clear, Count Keep

Measurement Mode

Single, Endless, Continuous

Display

Status, Error, Error Rate, Error Count, SyncLoss Count, Measurement Bit Number

Polarity Reversal Function

Supports polarity reversal for Data, Clock, Enable

Measured Result Reset Function

At BER measurement, hold sync status, clears measured value and measures from 0