Atec Agilent-N5183A User Manual
Page 6
6
Resolution 0.01
dB
Step attenuator
0 to 115 dB in 5 dB steps
(Option 1E1)
Amplitude
–15 to maximum specified output power with step attenuator
hold range
in 0 dB. Can be offset using Option 1E1 mechanical attenuator.
Amplitude switching speed
1, 2
Type
SCPI mode
2 ms (typ)
List/Step sweep mode
2 ms (typ)
Absolute level accuracy [dB]
3, 4
Frequency range
–20 to < –10 dBm
–10 to +10 dBm
> +10 dBm
250 kHz to 2 GHz
±1.4
±0.6
±0.6
2 to 20 GHz
±1.3
±0.9
±0.9
20 to 40 GHz
±1.3
±0.9
±1.0
Absolute level accuracy with Option 1E1 (dB)
4
Frequency range
–90 to < –75 dBm –75 to < –10 dBm –10 to +10 dBm > +10 dBm
250 kHz to 2 GHz
±1.4
±0.7
±0.6
±0.6
2 to 20 GHz
±1.6
±1.0
±0.9
±0.9
20 to 40 GHz
±2.0
±1.1
±0.9
±1.0
1.
Time from receipt of SCPI command or trigger signal to amplitude settled within 0.2 dB.
2.
Specification does not apply when switching from and to amplitudes where ALC levels are < –5 dBm
for Option 540 or < 0 dBm for Option 520.
3.
Level accuracy applies from –20 dBm to maximum output power between 15 °C and 35 °C.
4.
For temperatures outside this range, absolute level accuracy degrades by 0.01 dB/degree C for
frequencies ≤ 4.5 GHz and 0.02 dB/degree C for frequencies > 4.5 GHz.
For instruments with Type-N
connectors (Option 1ED), specifications are degraded typically 0.2 dB above 18 GHz. Specifications
do not apply above the maximum specified power.
Measured level accuracy Options 520 & 540 at –90 dBm
–1
–0.8
–0.6
–0.4
–0.2
0
0.2
0.4
0.6
0.8
1
0
5000
10000
15000
20000
25000
30000
35000
40000
Frequency (MHz)
Level accuracy (dB)
mean
+STD Dev
–STD Dev