6

Resolution 0.01

dB

Step attenuator

0 to 115 dB in 5 dB steps

(Option 1E1)

Amplitude

–15 to maximum specified output power with step attenuator

hold range

in 0 dB. Can be offset using Option 1E1 mechanical attenuator.

Amplitude switching speed

1, 2

Type
SCPI mode

2 ms (typ)

List/Step sweep mode

2 ms (typ)

Absolute level accuracy [dB]

3, 4

Frequency range

–20 to < –10 dBm

–10 to +10 dBm

> +10 dBm

250 kHz to 2 GHz

±1.4

±0.6

±0.6

2 to 20 GHz

±1.3

±0.9

±0.9

20 to 40 GHz

±1.3

±0.9

±1.0

Absolute level accuracy with Option 1E1 (dB)

4

Frequency range

–90 to < –75 dBm –75 to < –10 dBm –10 to +10 dBm > +10 dBm

250 kHz to 2 GHz

±1.4

±0.7

±0.6

±0.6

2 to 20 GHz

±1.6

±1.0

±0.9

±0.9

20 to 40 GHz

±2.0

±1.1

±0.9

±1.0

1.

Time from receipt of SCPI command or trigger signal to amplitude settled within 0.2 dB.

2.

Specification does not apply when switching from and to amplitudes where ALC levels are < –5 dBm

for Option 540 or < 0 dBm for Option 520.

3.

Level accuracy applies from –20 dBm to maximum output power between 15 °C and 35 °C.

4.

For temperatures outside this range, absolute level accuracy degrades by 0.01 dB/degree C for

frequencies ≤ 4.5 GHz and 0.02 dB/degree C for frequencies > 4.5 GHz.

For instruments with Type-N

connectors (Option 1ED), specifications are degraded typically 0.2 dB above 18 GHz. Specifications

do not apply above the maximum specified power.

Measured level accuracy Options 520 & 540 at –90 dBm

–1

–0.8

–0.6

–0.4

–0.2

0

0.2

0.4

0.6

0.8

1

0

5000

10000

15000

20000

25000

30000

35000

40000

Frequency (MHz)

Level accuracy (dB)

mean
+STD Dev
–STD Dev