beautypg.com

Atec Agilent-N5183A User Manual

Page 16

background image

16

Security (Option 006)

Memory sanitizing, memory sanitizing on power

on, and display blanking.

Self test

Internal diagnostic routines test most modules

in a preset condition. For each module, if its

node voltages are within acceptable limits, the

module “passes” the test.

Weight

≤ 13.8 kg (30 lb.) net,

≤ 28.4 kg (62 lb.) shipping

Dimensions

103 mm H x 426 mm W x 432 mm L

(4.07 in H x 16.8 in W x 17 in L)

Recommended calibration cycle 24

months

ISO compliant

The Agilent N5181A MXG is manufactured in an

ISO-9001 registered facility in concurrence with

Agilent Technologies’ commitment to quality.

Front panel connectors

1

RF output

Output impedance 50 Ω (nom)

Option 520

Precision APC-3.5 male, or Type-N with

Option

1ED

Option 532/540

Precision 2.4 mm male; plus 2.4 – 2.4 mm and

2.4 – 2.9 mm female adapters

Maximum reverse power

0.5 W, 0 Vdc

USB 2.0

Used with a memory stick for transferring

instrument states, licenses and other files into

or out of the instrument. Also used with U2000

series USB average power sensors. For a

current list of supported memory sticks, visit

www.agilent.com/find/MXG, click on Technical

Support, and refer to FAQs: WaveformDownloads

and

Storage.

Rear panel connectors

1

RF output

Outputs the RF signal via a precision N type

(Option 1EM)

female connector.

Sweep out

Generates output voltage, 0 to +10 V when the

signal generator is sweeping. This output can

also be programmed to indicate when the source

is settled or output pulse video and is TTL and

CMOS compatible in this mode. Output

impedance < 1 Ω, can drive 2k Ω. Damage

levels are ±15 V.

AM

External AM input. Nominal input impedance is

50 Ω. Damage levels are ±5 V.

FM

External FM input. Nominal input impedance is

50 Ω. Damage levels are ±5 V.

Pulse

External pulse modulation input. This input is

TTL and CMOS compatible. Low logic levels are

0 V and high logic levels are +1 V. Nominal input

impedance is 50 Ω. Input damage levels are

≤ –0.3 V and ≥ +5.3 V.

1.

All connectors are BNC unless otherwise noted.