Specification (spec):

Represents warranted performance of a calibrated instrument

that has been stored for a minimum of 2 hours within the operating temperature range
of 0 to 55 °C, unless otherwise stated, and after a 45 minute warm-up period. The
specifications include measurement uncertainty. Data represented in this document
are specifications unless otherwise noted.

Typical (typ):

Represents characteristic performance, which 80% of the instruments

manufactured will meet. This data is not warranted, does not include measurement
uncertainty, and is valid only at room temperature (approximately 25 °C).

Nominal (nom):

The expected mean or average performance, or an attribute whose

performance is by design, such as the 50 Ω connector. This data is not warranted and
is measured at room temperature (approximately 25 °C).

Measured (meas):

An attribute measured during the design phase for purposes of

communicating expected performance, such as amplitude drift vs. time. This data is
not warranted and is measured at room temperature (approximately 25 °C).

Note: All graphs contain measured data from several units at room temperature unless
otherwise noted.