14

Internal analog modulation source

(Option UNT)
Waveform

Sine

Rate range

0.1 Hz to 2 MHz (tunable to 3 MHz)

Resolution

0.1 Hz

Frequency accuracy

Same as RF reference source (nom)

External modulation inputs

(Requires Option UNT)

Modulation types

FM, AM, phase mod, pulse mod

Input impedance

50 Ω (nom)

Simultaneous modulation

1

All modulation types (FM, AM, φM and pulse modulation) may be simultaneously

enabled except: FM and phase modulation can not be combined; two modulation

types

can not be simultaneously generated using the same modulation source. For example,
AM and FM can run concurrently and will modulate the output RF. This is useful for
simulating signal impairments.

1.

If AM or pulse modulation are on, then phase and FM specifications do not apply.