1. Specifications apply in CW and list/step sweep modes over the 15 to 35 ºC temperature range, with attenuator hold off
(normal operating mode). Degradation outside this range, for ALC power levels > –10 dBm, is typically < 0.3 dB. In ramp sweep
mode (with Option 007), specifications are typical. For instruments with type-N connectors (Option 1ED), specifications are

degraded typically 0.2 dB above 18 GHz. Specifications do not apply above the maximum specified power.
2. Options 550 and 567: 0.03dB/°C (typ) above 2 GHz.
3. Compatible with Agilent EPM Series (E4418B and E4419B) power meters.

Level accuracy with step attenuator (Option 1E1)

1

(dB)

Frequency

> +10 dBm +10 to 0 dBm 0 to –10 dBm –10 to –70 dBm –70 to –90 dBm

250 kHz to 2 GHz ±0.6

±0.6

±0.6

±0.7

±0.8

> 2 to 20 GHz

±0.8

±0.8

±0.8

±0.9

±1.0

> 20 to 40 GHz ±1.0

±0.9

±0.9

±1.0

±2.0

> 40 to 50 GHz

---

±1.3

±0.9

±1.5

±2.5

> 50 to 67 GHz

---

±1.5

±1.0

±1.5 (typ)

±2.5 (typ)

Level accuracy (measured)

Resolution

0.01 dB

Temperature stability

0.01 dB/°C (typ)

2

User flatness correction
Number of points

2 to 1601 points/table

Number of tables

Up to 10,000, memory limited

Path loss

Arbitrary, within attenuator range

Entry modes

Remote power meter

3

, remote bus, manual

(user

edit/view)

0.25

0.2

0.15

0.1

0.05

0

-0.05

-0.1

-0.15

-0.2

Errorr (dB)

Frequency (GHz)

Option 520 with Option 1E1 at -110 dBm (me

0

4

8

12

16

20

Option 550/567 with Option 1E1 at -90 dBm

Errorr (dB)

Frequency (GHz)

0

10

20

30

40

50

60

70

1

0.5

0

-0.5

-1

0.8

0.6

0.4

0.2

0

-0.2

-0.4

-0.6

-0.8

Errorr (dB)

Frequency (GHz)

Option 532/540

with Option 1E1 at -110 dBm

0

10

20

30

40