Measured phase noise with E5500 and plotted without spurs
Standard SSB phase noise (measured)
1
-20
-30
-40
-50
-60
-70
-80
-90
-100
-110
-120
-130
-140
-150
-160
-170
z
H
G
7
6
40 GHz
20 GHz
10 GHz
1 GHz
L(f) (dBc/Hz) vs. f (Hz)
10
100
1 k
10 k
100 k
1 M
10 M 100 M
Option UNX absolute SSB phase noise (measured)
-20
-30
-40
-50
-60
-70
-80
-90
-100
-110
-120
-130
-140
-150
-160
-170
L(f) (dBc/Hz) vs. f (Hz)
1
10
100
1 k
10 k 100 k 1 M
10 M 100 M
z
H
G
7
6
40 GHz
20 GHz
10 GHz
1 GHz
L(f) (dBc/Hz) vs. f (Hz)
Option UNX residual phase noise (measured)
-20
-30
-40
-50
-60
-70
-80
-90
-100
-110
-120
-130
-140
-150
-160
-170
1
10
100
1 k
10 k 100 k 1 M
10 M 100 M
10 GHz
3 GHz
1 GHz
L(f) (dBc/Hz) vs. f (Hz)
AM noise at 10 GHz (measured)
-20
-30
-40
-50
-60
-70
-80
-90
-100
-110
-120
-130
-140
-150
-160
-170
1
10
100
1 k
10 k
100 k
1 M
10 M 100 M
AM noise at 10 GHz
Residual phase noise
Option UNX phase noise
Standard phase noise
1.
Measured standard performance applies to units with serial numbers ending with 48050000 or greater. For units with lower
serial numbers, refer to the data sheet shipped with the unit or the version of this document dated November 5, 2007.