beautypg.com

General specifications – Atec Agilent-E8257D User Manual

Page 19

background image

19

1. Storage below –20 °C instrument states may be lost.
2. As is the case with all signal generation equipment, phase noise specifications are not warranted in a vibrating environment.

Power requirements

90 to 132 VAC 47 to 64 Hz or 365 to 435 Hz; or

195 to 267 VAC 47 to 64 Hz, (automatically selected);

< 250 W typical, 300 W maximum.

Operating temperature range

0 to 55 °C

Storage temperature range

1

–40 to 70 °C

Altitude <

4,572 m (15,000 ft.)

Environmental testing

Samples of this product have been tested in

accordance with the Agilent Environmental

Test Manual and verified to be robust against

the environmental stresses of storage,

transportation, and end-use; those stresses

include but are not limited to temperature,

humidity, shock, vibration, altitude, and power

line conditions. Test methods are aligned

with IEC 60068-2 and levels are similar to

MIL-PRF-28800F Class 3.

2

EMC

Meets the conducted and radiated interference

and immunity requirements of IEC/EN 61326-1.

Meets radiated emission requirements of CISPR

Pub 11/1997 Group 1 class A.

Storage registers

Memory is shared by instrument states and

sweep list files. There is 14 MB of flash memory

available in the E8257D PSG. Depending on

how the memory is used, a maximum of 1000

instrument states can be saved.

Security Display

blanking

Memory clearing functions

(see Application Note Security of Agilent Signal

Generators Issues and Solutions, literature

number

5989-1091EN)

Compatibility

Agilent 83550 Series Millimeter Heads and OML

millimeter source modules.

Agilent 8757D scalar network analyzers.

Agilent EPM Series power meters.

Self-test

Internal diagnostic routine tests most modules

(including microcircuits) in a preset condition.

For each module, if its node voltages are within

acceptable limits, then the module “passes”

the

test.

Weight

< 22 kg (48 lb.) net, < 30 kg (68 lb.) shipping

Dimensions

178 mm H x 426 mm W x 515 mm D

(7” H x 16.8” W x 20.3” D in.)

Recommended calibration cycle 24