Maximum deviation

2

Frequency

Maximum deviation

250 kHz to 250 MHz

2 MHz

> 250 to 500 MHz

1 MHz

> 500 MHz to 1 GHz

2 MHz

> 1 GHz to 2 GHz

4 MHz

> 2 GHz to 3.2 GHz

8 MHz

> 3.2 GHz to 10 GHz

16 MHz

> 10 GHz to 20 GHz

32 MHz

> 20 GHz to 40 GHz

64 MHz

> 40 GHz to 67 GHz

128 MHz

Resolution

0.1% of deviation or 1 Hz, whichever is greater

Deviation accuracy

< ± 3.5% of FM deviation + 20 Hz

(1 kHz rate, deviations < N x 800 kHz)

Modulation frequency response

3

(at 100 kHz deviation)

Path [coupling]

1 dB bandwidth

3 dB bandwidth (typ)

FM path 1 [DC]

DC to 100 kHz

DC to 10 MHz

FM path 2 [DC]

DC to 100 kHz

DC to 1 MHz

FM path 1 [AC]

20 Hz to 100 kHz

5 Hz to 10 MHz

FM path 2 [AC]

20 Hz to 100 kHz

5 Hz to 1 MHz

DC FM

4

carrier offset

±0.1% of set deviation + (N x 8 Hz)

Distortion

< 1% (1 kHz rate, deviations < N x 800 kHz)

Sensitivity

±1 V

peak

for indicated deviation

Paths

FM1 and FM2 are summed internally for composite

modulation. Either path may be switched to any one of

the modulation sources: Ext1, Ext2, internal1, internal2.

The FM2 path is limited to a maximum rate of 1 MHz.

The FM2 path must be set to a deviation less than FM1.

Maximum deviation

6

Frequency

Normal BW mode

High BW mode

250 kHz to 250 MHz

20 rad

2 rad

> 250 to 500 MHz

10 rad

1 rad

> 500 MHz to 1 GHz

20 rad

2 rad

> 1 GHz to 2 GHz

40 rad

4 rad

> 2 GHz to 3.2 GHz

80 rad

8 rad

> 3.2 GHz to 10 GHz

160 rad

16 rad

> 10 GHz to 20 GHz

320 rad

32 rad

> 20 GHz to 40 GHz

640 rad

64 rad

> 40 GHz to 67 GHz

1280 rad

128 rad

Resolution

0.1% of set deviation

Deviation accuracy

< ±5% of deviation + 0.01 radians (1 kHz rate, normal

BW

mode)

Modulation frequency response

7

Normal BW mode

High BW mode

Rates (3 dB BW)

DC to 100 kHz

DC to 1 MHz (typ)

8

Distortion

< 1 % (1 kHz rate, Total Harmonic Distortion (THD),

dev < N x 80 rad, normal BW mode)

Sensitivity

±1 V

peak

for indicated deviation

Paths

FM1 and FM2 are summed internally for composite

modulation. Either path may be switched to any one of

the modulation sources: Ext1, Ext2, internal1, internal2.

The

FM2 path must be set to a deviation less than FM1.

1. Above 50 GHz, FM is useable; however performance is not warranted.
2. Through any combination of path1, path2, or path1 + path2.
3. Specifications apply in CW and list/step sweep modes. During ramp sweep operation (Option 007), 3 dB bandwidth is typically 50 kHz to 10 MHz
(FM1 path), and 50 kHz to 1 MHz (FM2 path).
4. At the calibrated deviation and carrier frequency, within 5 °C of ambient temperature at time of user calibration.
5. Above 50 GHz, phase modulation is useable; however performance is not warranted.
6. Through any combination of path1, path2, or path1 + path2.
7. Specifications apply in CW and list/step sweep modes. During ramp sweep operation (Option 007), 3 dB bandwidth is typically 50 kHz to 1 MHz
(high BW mode).
8. Path 1 is useable to 4 MHz for external inputs less than 0.3 V peak.

Frequency modulation

1

(Option UNT)

Phase modulation

5

(Option UNT)