Frequency modulation, Option unt) phase modulation, Option unt) – Atec Agilent-E8257D User Manual
Page 13
13
Maximum deviation
2
Frequency
Maximum deviation
250 kHz to 250 MHz
2 MHz
> 250 to 500 MHz
1 MHz
> 500 MHz to 1 GHz
2 MHz
> 1 GHz to 2 GHz
4 MHz
> 2 GHz to 3.2 GHz
8 MHz
> 3.2 GHz to 10 GHz
16 MHz
> 10 GHz to 20 GHz
32 MHz
> 20 GHz to 40 GHz
64 MHz
> 40 GHz to 67 GHz
128 MHz
Resolution
0.1% of deviation or 1 Hz, whichever is greater
Deviation accuracy
< ± 3.5% of FM deviation + 20 Hz
(1 kHz rate, deviations < N x 800 kHz)
Modulation frequency response
3
(at 100 kHz deviation)
Path [coupling]
1 dB bandwidth
3 dB bandwidth (typ)
FM path 1 [DC]
DC to 100 kHz
DC to 10 MHz
FM path 2 [DC]
DC to 100 kHz
DC to 1 MHz
FM path 1 [AC]
20 Hz to 100 kHz
5 Hz to 10 MHz
FM path 2 [AC]
20 Hz to 100 kHz
5 Hz to 1 MHz
DC FM
4
carrier offset
±0.1% of set deviation + (N x 8 Hz)
Distortion
< 1% (1 kHz rate, deviations < N x 800 kHz)
Sensitivity
±1 V
peak
for indicated deviation
Paths
FM1 and FM2 are summed internally for composite
modulation. Either path may be switched to any one of
the modulation sources: Ext1, Ext2, internal1, internal2.
The FM2 path is limited to a maximum rate of 1 MHz.
The FM2 path must be set to a deviation less than FM1.
Maximum deviation
6
Frequency
Normal BW mode
High BW mode
250 kHz to 250 MHz
20 rad
2 rad
> 250 to 500 MHz
10 rad
1 rad
> 500 MHz to 1 GHz
20 rad
2 rad
> 1 GHz to 2 GHz
40 rad
4 rad
> 2 GHz to 3.2 GHz
80 rad
8 rad
> 3.2 GHz to 10 GHz
160 rad
16 rad
> 10 GHz to 20 GHz
320 rad
32 rad
> 20 GHz to 40 GHz
640 rad
64 rad
> 40 GHz to 67 GHz
1280 rad
128 rad
Resolution
0.1% of set deviation
Deviation accuracy
< ±5% of deviation + 0.01 radians (1 kHz rate, normal
BW
mode)
Modulation frequency response
7
Normal BW mode
High BW mode
Rates (3 dB BW)
DC to 100 kHz
DC to 1 MHz (typ)
8
Distortion
< 1 % (1 kHz rate, Total Harmonic Distortion (THD),
dev < N x 80 rad, normal BW mode)
Sensitivity
±1 V
peak
for indicated deviation
Paths
FM1 and FM2 are summed internally for composite
modulation. Either path may be switched to any one of
the modulation sources: Ext1, Ext2, internal1, internal2.
The
FM2 path must be set to a deviation less than FM1.
1. Above 50 GHz, FM is useable; however performance is not warranted.
2. Through any combination of path1, path2, or path1 + path2.
3. Specifications apply in CW and list/step sweep modes. During ramp sweep operation (Option 007), 3 dB bandwidth is typically 50 kHz to 10 MHz
(FM1 path), and 50 kHz to 1 MHz (FM2 path).
4. At the calibrated deviation and carrier frequency, within 5 °C of ambient temperature at time of user calibration.
5. Above 50 GHz, phase modulation is useable; however performance is not warranted.
6. Through any combination of path1, path2, or path1 + path2.
7. Specifications apply in CW and list/step sweep modes. During ramp sweep operation (Option 007), 3 dB bandwidth is typically 50 kHz to 1 MHz
(high BW mode).
8. Path 1 is useable to 4 MHz for external inputs less than 0.3 V peak.
Frequency modulation
1
(Option UNT)
Phase modulation
5
(Option UNT)