beautypg.com

Atec Agilent-E8257D User Manual

Page 15

background image

15

Modulation types

AM, FM, and

FM

Input impedance

50 or 600

(nom) switched

High/low indicator
(100 Hz to 10 MHz BW,

Activated when input level error exceeds 3% (nom)

ac coupled inputs only)

Dual function generators provides two independent signals (internal1 and internal2) for
use with AM, FM,

FM, or LF Out.

Waveforms

Sine, square, positive ramp, negative ramp, triangle,

Gaussian noise, uniform noise, swept sine, dual sine

1

Rate range
Sine

0.5 Hz to 1 MHz

Square, ramp, triangle

0.5 Hz to 100 kHz

Resolution

0.5 Hz

Accuracy

Same as timebase

LF Out
Output

Internal1 or internal2. Also provides monitoring of

internal1or internal2 when used for AM, FM, or

FM.

Amplitude

0 to 3 V

peak

, (nom) into 50 Ω

Output impedance

50 Ω (nom)

Swept sine mode: (frequency, phase continuous)
Operating modes

Triggered or continuous sweeps

Frequency range

1 Hz to 1 MHz

Sweep rate

0.5 Hz to 100 kHz sweeps/s, equivalent to sweep times

10 us to 2 s

Resolution

0.5 Hz (0.5 sweep/s)

External modulation inputs

(Ext1 & Ext2)
(Option UNT)

Internal modulation source

(Option UNT)

1. Internal2 is not available when using swept sine or dual sine modes.