Step attenuator

1

(Option 1E1)

Options 520, 532, and 540

0 dB and 5 dB to 115 dB in 10 dB steps

Options 550 and 567

0 dB to 90 dB in 10 dB steps

Maximum available power (measured)

Attenuator hold range
Minimum

From –20 dBm to maximum specified output power with step

attenuator in 0 dB position. Can be offset using Option 1E1 attenuator.

Amplitude switching speed

2

ALC on or off

< 3 ms (typ)

(without power search)
Level accuracy

3

(dB)

Frequency

> +10 dBm

+10 to 0 dBm

0 to –10 dBm

–10 to –20 dBm

250 kHz to 2 GHz

±0.6

±0.6

±0.6

±1.4

> 2 GHz to 20 GHz

±0.8

±0.8

±0.8

±1.2

> 20 to 40 GHz

±1.0

±0.9

±0.9

±1.3

> 40 to 50 GHz

---

±1.3

±0.9

±1.2

> 50 to 67 GHz

---

±1.5

±1.0

±1.2 (typ)

6

Option 520 with Option 1EA (measured

26

24

22

20

18

16

0 5000

Frequency (MHz)

Option 550/567 with Option 1EA (measur

Output power (dBm)

Frequency (GHz)

30

25

20

15

10

5

0

0

10

20

30

40

50

60

70

Option 532/540 with Option 1EA (measu

26

24

22

20

18

16

0 10000

Frequency (MHz)

1. The step attenuator provides coarse power attenuation to achieve low power levels. Fine power level adjustment is provided
by the ALC (Automatic Level Control) within the attenuator hold range.
2. To within 0.1 dB of final amplitude within one attenuator range. Add 10 to 50 ms when using power search.
3. Specifications apply in CW and list/step sweep modes over the 15 to 35 ºC temperature range with the ALC on. Degradation
outside this range, for power levels > –10 dBm, is typically < 0.3 dB. In ramp sweep mode (with Option 007), specifications
are typical. For instruments with Type-N connectors (Option 1ED), specifications are degraded typically 0.2 dB above 18 GHz.