Output – Atec Agilent-E8257D User Manual
Page 5
5
Power
1
(dBm)
Frequency range
Standard
Option 1EA
spec. (typ)
Option 520:
250 kHz to 3.2 GHz
–20 to +15
5
–20 to +16 (+19)
250 kHz to 3.2 GHz with Option UNW
–20 to +11
–20 to +11 (+14)
250 kHz to 3.2 GHz with Option 1EH
–20 to +13
2
–20 to +13 (+16)
2
250 kHz to 3.2 GHz with Options UNW and 1EH –20 to +10
2
–20 to +10 (+13)
2
> 3.2 Ghz to 5.2 GHz
–20 to +15
5
–20 to +22 (+23)
4
> 5.2 Ghz to 12 GHz
–20 to +15
5
–20 to +23 (+24)
4
> 12 Ghz to 20 GHz
–20 to +15
5
–20 to +21 (+23)
4
Options 532 and 540:
250 kHz to 3.2 GHz
–20 to +11
5
–20 to +15 (+18)
250 kHz to 3.2 GHz with Option UNW
–20 to +9
–20 to +10 (+13)
250 kHz to 3.2 GHz with Option 1EH
–20 to +9
–20 to +12 (+15)
2
250 kHz to 3.2 GHz with Options UNW and 1EH –20 to +9
2
–20 to +9 (+12)
2
> 3.2 to 17 GHz
–20 to +11
5
–20 to +19 (+21)
4
> 17 to 37 GHz
–20 to +11
5
–20 to +16 (+19)
4
> 37 to 40 GHz
–20 to +11
5
–20 to +14 (+17)
Options 550 and 567:
250 kHz to 3.2 GHz
–20 to +5
–20 to +14 (+17)
250 kHz to 3.2 GHz with Option UNW
–20 to +5
–20 to +9 (+12)
250 kHz to 3.2 GHz with Option 1EH
–20 to +5
–20 to +11 (+14)
2
250 kHz to 3.2 GHz with Options UNW and 1EH –20 to +5
–20 to +8 (+11)
2
> 3.2 to 10 GHz
–20 to +5
–20 to +14 (+21)
> 10 to 20 GHz
–20 to +5
–20 to +14 (+17)
> 20 to 30 GHz
–20 to +5
–20 to +11 (+17)
> 30 to 65 GHz
–20 to +5
–20 to +11 (+14)
> 65 to 67 GHz
–20 to +5
–20 to +10 (+14)
> 67 to 70 GHz
–20 to +5 (typ)
–20 to +8 (typ)
Option 520 with step attenuator (Option 1E1):
250 kHz to 3.2 GHz
–135 to +13
5
–135 to +15 (+18)
250 kHz to 3.2 GHz with Option UNW
–135 to +10
–135 to +10 (+13)
250 kHz to 3.2 GHz with Option 1EH
–135 to +11
3
–135 to +12 (+15)
2
250 kHz to 3.2 GHz with Options UNW and 1EH –135 to +9
2
–135 to +9 (+12)
2
> 3.2 GHz to 10 GHz
–135 to +13
5
–135 to +21 (+22)
4
> 10 GHz to 20 GHz
–135 to +13
5
–135 to +19 (+20)
4
Options 532 and 540 with step attenuator (Option 1E1):
250 kHz to 3.2 GHz
–135 to +9
5
–135 to +14 (+17)
250 kHz to 3.2 GHz with Option UNW
–135 to +7
–135 to +9 (+12)
250 kHz to 3.2 GHz with Option 1EH
–135 to +7
–135 to +11 (+14)
2
250 kHz to 3.2 GHz with Options UNW and 1EH –135 to +7
3
–135 to +8 (+11)
2
> 3.2 to 17 GHz
–135 to +9
5
–135 to +17 (+20)
4
> 17 to 37 GHz
–135 to +9
5
–135 to +14 (+17)
4
> 37 to 40 GHz
–135 to +9
5
–135 to +12 (+16)
Options 550 and 567 with step attenuator (Option 1E1):
250 kHz to 3.2 GHz
–110 to +3
–110 to +13 (+16)
250 kHz to 3.2 GHz with Option UNW
–110 to +3
–110 to +8 (+11)
250 kHz to 3.2 GHz with Option 1EH
–110 to +3
–110 to +10 (+13)
2
250 kHz to 3.2 GHz with Options UNW and 1EH –110 to +3
–110 to +7 (+10)
2
> 3.2 to 10 GHz
–110 to +3
–110 to +13 (+20)
> 10 to 20 GHz
–110 to +3
–110 to +13 (+16)
> 20 to 30 GHz
–110 to +3
–110 to +9 (+16)
> 30 to 65 GHz
–110 to +3
–110 to +9 (+12)
> 65 to 67 GHz
–110 to +3
–110 to +8 (+12)
> 67 to 70 GHz
–110 to +3 (typ) –110 to +6 (typ)
Output
1. Maximum power specifications are warranted from 15 to 35 °C, and is typical from 0 to 15 °C. Maximum power over the 35 to 55 °C
range typically degrades less than 2 dB.
2. With harmonic filters switched off. With filters on, maximum output power is reduced 3 dB for frequencies below 2 GHz.
3. With harmonic filters switched off. With filters on, maximum output power is reduced 2 dB for frequencies below 2 GHz.
4. Specification applies to units with serial numbers ending with 45470000 or greater. For units with lower serial numbers, refer to the data
sheet shipped with the unit or the version of this document dated December 16, 2004.
5. Standard performance applies to units with serial numbers ending with 48050000 or greater. For units with lower serial
numbers, refer to the data sheet shipped with the unit or the version of this document dated November 5, 2007.