beautypg.com

Infiniivision x-series physical characteristics, Connectivity, General and environmental characteristics – Atec Agilent-2000-X Series User Manual

Page 20: Nonvolatile storage, Included standard with oscilloscope

background image

20

Connectivity

Standard ports

One USB 2.0 hi-speed device port on rear panel. Supports USBTMC protocol.
Two USB 2.0 hi-speed host ports, front and rear panel
Supports memory devices and printers

Optional ports

GPIB, LAN, WVGA video out

General and environmental characteristics

Power line consumption

100 watts

Power voltage range

100-120V, 50/60/400 Hz; 100-240V, 50/60 Hz ±10% auto ranging

Temperature

Operating: 0 to +55 °C
Non-operating: –30 to +71 °C

Humidity

Operating: Up to 80% RH at or below +40 °C; up to 45% RH up to +50 °C
Non-operating: Up to 95% RH up to 40 °C; up to 45% RH up to 50 °C

Altitude

Operating: up to 4,000 m, Non-operating 15,300 m

Electromagnetic compatibility

Meets EMC Directive (2004/108/EC), meets or exceeds IEC 61326-1:2005/EN
61326-1:2006 Group 1 Class A requirement
CISPR 11/EN 55011
IEC 61000-4-2/EN 61000-4-2
IEC 61000-4-3/EN 61000-4-3
IEC 61000-4-4/EN 61000-4-4
IEC 61000-4-5/EN 61000-4-5
IEC 61000-4-6/EN 61000-4-6
IEC 61000-4-11/EN 61000-4-11
Canada: ICES-001:2004
Australia/New Zealand: AS/NZS

Safety

UL61010-1 2nd edition, CAN/CSA22.2 No. 61010-1-04

Dimensions

381 mm (15 in) W x 204 mm (8 in) H x 142 mm (5.6 in) D

Weight

Net: 3.9 kg (8.5 lbs), shipping: 4.1 kg (9.0 lbs)

Nonvolatile storage

Reference waveform display

2 internal waveforms or USB thumb drive

Waveform storage

Set up, .bmp, .png, .csv, ASCII, XY, reference waveforms, .alb, .bin, lister, mask , HDFS

Max USB flash drive size

Supports industry standard flash drives

Set ups without USB flash drive

10 internal setups

Set ups with USB flash drive

Limited by size of USB drive

Oscilloscopes redefined: Breakthrough technology delivers more scope for the same budget

InfiniiVision X-Series physical characteristics

Included standard with oscilloscope

Standard 5-year warranty*

Standard secure erase

Standard Probe

N2862B 150 MHz 10:1 passive probe

Standard one per channel for 70 and 100 MHz models

N2863B 300 MHz, 10:1 passive probe

Standard one per channel for 200 MHz models

N6459-60001 8-channel logic probe and accessory kit

Standard on MSO models or with DSOX2MSO upgrade

Built-in help language support

for English, Japanese, simplified Chinese, traditional Chinese, Korean, German, French, Spanish,

Russian, Portuguese and Italian, Certificate of Calibration, Documentation CD

Interface language support GUI menus:

English, Japanese, simplified Chinese, traditional Chinese, Korean, German, French,

Spanish, Russian, Portuguese, Thai, Polish and Italian

Localized power cord

*Applies to all orders on or after 1/1/2013.

For MET/CAL procedures, click on the Cal Labs solutions link below

Cal Labs Solutions

http://www.callabsolutions.com/MetCALandCLS.asp

These procedures are FREE to customers.