Waveform math

Operators

Add, subtract, multiply, FFT

FFT

Windows: Hanning, flat top, rectangular; Blackman-Harris - up to 64 kpts resolution

Sources

Math functions available between any two channels

Display characteristics

Display

8.5-inch WVGA

Resolution

800 (H) x 480 (V) pixel format (screen area)

Interpolation

Sin(x)/x interpolation (using FIR filter; used when there is less than one sample per column of the display)

Persistence

Off, infinite, variable persistence (100ms-60s)

Intensity gradation

64 intensity levels

Modes

Normal
XY - XY mode changes the display from voltage versus time scale to a volts versus volts scale
Roll – Displays the waveform moving across the screen from right to left much like a strip chart recorder

MSO (digital channels)

Upgradable from DSO

Yes

MSO channels

8 channels (D0 to D7)

Maximum sample rate

1 GSa/s

Maximum record length

50 kpts per channel (digital channels only)
12.5 kpts per channel (analog and digital channels)

Threshold selections

TTL (+1.4 V), CMOS (+2.5 V), ECL (-1.3 V), User-definable (±8.0 V in 10 mV stops)

Threshold accuracy*

± (100 mV + 3% of threshold settings)

Maximum input dynamic
range

±10 V about threshold

Minimum voltage swing

500 mVpp

Input impedance

100 kΩ ± 2% at probe tip, ~8 pF

Minimum detectable pulse
width

5 ns

Channel-to-channel skew

2 ns (typical), 3 ns (maximum)

Oscilloscopes redefined: Breakthrough technology delivers more scope for the same budget

Performance characterisitics

* Denotes warranted specifications, all others are typical.
Specifications are valid after a 30-minute warm-up period and from ±10 °C firmware calibration temperature.
** 1 mV/div and 2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for
2 mV/div sensitivity setting.

Cursors

Types

Amplitude, time , frequency (FFT), manual, tracking, binary, HEX

Measurements

ΔT, 1/ΔT, ΔV/X, 1/ΔX, ΔY, Phase and Ratio

Cursors**

• Single cursor accuracy:
±[DC vertical gain accuracy + DC vertical offset accuracy + 0.25% full scale]
• Dual cursor accuracy:
±[DC vertical gain accuracy + 0.5% full scale]*

Automatic waveforms measurements

Voltage

Snapshot all, maximum, minimum, peak-to-peak, top, base, amplitude, overshoot, preshoot, average- N
cycles, average-full screen, DC RMS- N cycles, DC RMS- full screen, AC RMS- N cycles

Time

Period, frequency, rise time, fall time, + width, – width, duty cycle, delay A→B (rising edge), delay A→B
(falling edge), phase A→B (rising edge,) and phase A→B (falling edge)