Specification overview
2002A
2004A
2012A
2014A
2022A
2024A
Bandwidth * (–3 dB)
70 MHz
100 MHz
200 MHz
“Calculated rise time (10-90%)”
≤ 5 ns
≤ 3.5 ns
≤ 1.75 ns
Input channels
DSOX
2
4
2
4
2
4
MSOX
2 + 8
4 + 8
2 + 8
4 + 8
2 + 8
4 + 8
Maximum sample rate
2 GSa/s half-channel interleaved, 1 GSa/s per channel
Maximum memory depth
100 kpts per channel (standard), 1 Mpt per channel (optional with DSOX2MEMUP)
Display size and type
8.5-inch WVGA with 64 levels of intensity grading
Waveform update rate
> 50,000 waveforms per second
Vertical system analog channels
Input coupling
AC, DC
Input sensitivity range
1 mV/div to 5 V/div**
Input impedance
1 MΩ ± 2% (11 pF)
Vertical resolution
8 bits
Dynamic range
±8 divisions from center screen
Maximum input voltage
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
CAT II 300 Vrms, 400 Vpk; with N2862B or N2863B 10:1 probe: 300 Vrms
DC vertical accuracy
±[DC vertical gain accuracy + DC vertical offset accuracy + 0.25% full scale] **
DC vertical gain accuracy*
±3% full scale (≥ 10 mV/div); ±4% full scale (< 10 mV/div) **
DC vertical offset accuracy
±0.1 div ± 2mV ±1% of offset setting
Channel-to-channel isolation 200 MHz ~ 1 GHz ≥ 40 dB from DC to maximum specified bandwidth of each model
1.5 GHz
≥ 40 dB from DC to 1 GHz, ≥ 35 dB from 1 GHz to 1.5 GHz
Position/offset range
1 MΩ
1 mV to 200 mV/div: ±2 V, > 200 mV to 5 V/div: ± 50V
Hardware bandwidth limits
Approximately 20 MHz (selectable)
Horizontal system analog channels
2002A
2004A
2012A
2014A
2022A
2024A
Time base range
5 ns/div to 50 s/div
2 ns/div to 50 s/div
Horizontal resolution
2.5 ps
Time base accuracy*
25 ppm ±5 ppm per year (aging)
Time base delay time
range
Pre-trigger
Greater of 1 screen width or 200 μs (400 μs in interleaving mode)
Post-trigger
1 s to 500 s
Channel-to-channel deskew range
± 100 ns
Δ Time accuracy (using cursors)
± (time base accuracy* reading) ± (0.0016 * screen width) ± 100 ps