Oscilloscopes redefined: Breakthrough technology delivers more scope for the same budget

Performance characteristics

* Denotes warranted specifications, all others are typical.
Specifications are valid after a 30-minute warm-up period and from ±10 °C firmware calibration temperature.
** 1 mV/div and 2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 1 mV/div and 2 mV/div sensitivity setting.

Specification overview

2002A

2004A

2012A

2014A

2022A

2024A

Bandwidth * (–3 dB)

70 MHz

100 MHz

200 MHz

“Calculated rise time (10-90%)”

≤ 5 ns

≤ 3.5 ns

≤ 1.75 ns

Input channels

DSOX

2

4

2

4

2

4

MSOX

2 + 8

4 + 8

2 + 8

4 + 8

2 + 8

4 + 8

Maximum sample rate

2 GSa/s half-channel interleaved, 1 GSa/s per channel

Maximum memory depth

100 kpts per channel (standard), 1 Mpt per channel (optional with DSOX2MEMUP)

Display size and type

8.5-inch WVGA with 64 levels of intensity grading

Waveform update rate

> 50,000 waveforms per second

Vertical system analog channels

Input coupling

AC, DC

Input sensitivity range

1 mV/div to 5 V/div**

Input impedance

1 MΩ ± 2% (11 pF)

Vertical resolution

8 bits

Dynamic range

±8 divisions from center screen

Maximum input voltage

CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk

CAT II 300 Vrms, 400 Vpk; with N2862B or N2863B 10:1 probe: 300 Vrms

DC vertical accuracy

±[DC vertical gain accuracy + DC vertical offset accuracy + 0.25% full scale] **

DC vertical gain accuracy*

±3% full scale (≥ 10 mV/div); ±4% full scale (< 10 mV/div) **

DC vertical offset accuracy

±0.1 div ± 2mV ±1% of offset setting

Channel-to-channel isolation 200 MHz ~ 1 GHz ≥ 40 dB from DC to maximum specified bandwidth of each model

1.5 GHz

≥ 40 dB from DC to 1 GHz, ≥ 35 dB from 1 GHz to 1.5 GHz

Position/offset range

1 MΩ

1 mV to 200 mV/div: ±2 V, > 200 mV to 5 V/div: ± 50V

Hardware bandwidth limits

Approximately 20 MHz (selectable)

Horizontal system analog channels

2002A

2004A

2012A

2014A

2022A

2024A

Time base range

5 ns/div to 50 s/div

2 ns/div to 50 s/div

Horizontal resolution

2.5 ps

Time base accuracy*

25 ppm ±5 ppm per year (aging)

Time base delay time
range

Pre-trigger

Greater of 1 screen width or 200 μs (400 μs in interleaving mode)

Post-trigger

1 s to 500 s

Channel-to-channel deskew range

± 100 ns

Δ Time accuracy (using cursors)

± (time base accuracy* reading) ± (0.0016 * screen width) ± 100 ps