9

Uncorrected System Performance

User correction: OFF, System error correction: ON

Test Port Output (Source)

Specification (dB)

Description

30 to 300 kHz

300 kHz to 3 GHz

3 to 6 GHz

6 to 8.5 GHz

Directivity

15 (Typ.)

25

20

15

Source match

15 (Typ.)

25

20

15

Refl ection tracking

± 1.5 (Typ.)

± 1.0

± 1.5

± 1.5

Transmission tracking

± 1.5 (Typ.)

± 1.0

± 1.5

± 1.5

Test port output frequency

Description

Specification

Typical

Frequency range

1

Option 245
Option 285

30 kHz to 4.5 GHz
30 kHz to 8.5 GHz

Resolution

1 Hz

Source Stability
Standard
Option 1E5

± 7 ppm (5 to 40 ºC)
± 1 ppm (5 to 40 ºC)
± 0.5 ppm/year

CW accuracy
Standard
Option 1E5

± 7 ppm
± 1 ppm

Test port output power

2

Description

Specification

SPD

Nominal Power
(Preset Power)

0 dBm

Range

3

30 to 300 kHz
300 kHz to 3 GHz
3 to 6 GHz
6 to 8.5 GHz

–85 to 10 dBm
–85 to 16 dBm
–85 to 12 dBm
–85 to 10 dBm

Max Leveled Power
30 to 300 kHz
300 kHz to 1 GHz
1 to 3 GHz
3 to 8.5 GHz

16 dBm
20 dBm
18 dBm
Slope from 18 dBm
(3 GHz) to 12 dBm
(8.5 GHz)

Specification (dB)

Description

30 to 300 kHz

300 kHz to 100 MHz 100 MHz to 2 GHz

2 to 4.5 GHz

4.5 to 8.5 GHz

Load match

14 (Typ.)

22

15

11

8

1. Frequency can be set down to 9 kHz
2. Source output performance on port 1 only. Port 2 output performance is typical.
3. Power can be set from –109 to 20 dBm