Uncorrected system performance, Test port output (source) – Atec Agilent-N5072A Series User Manual
Page 9
9
Uncorrected System Performance
User correction: OFF, System error correction: ON
Test Port Output (Source)
Specification (dB)
Description
30 to 300 kHz
300 kHz to 3 GHz
3 to 6 GHz
6 to 8.5 GHz
Directivity
15 (Typ.)
25
20
15
Source match
15 (Typ.)
25
20
15
Refl ection tracking
± 1.5 (Typ.)
± 1.0
± 1.5
± 1.5
Transmission tracking
± 1.5 (Typ.)
± 1.0
± 1.5
± 1.5
Test port output frequency
Description
Specification
Typical
Frequency range
1
Option 245
Option 285
30 kHz to 4.5 GHz
30 kHz to 8.5 GHz
Resolution
1 Hz
Source Stability
Standard
Option 1E5
± 7 ppm (5 to 40 ºC)
± 1 ppm (5 to 40 ºC)
± 0.5 ppm/year
CW accuracy
Standard
Option 1E5
± 7 ppm
± 1 ppm
Test port output power
2
Description
Specification
SPD
Nominal Power
(Preset Power)
0 dBm
Range
3
30 to 300 kHz
300 kHz to 3 GHz
3 to 6 GHz
6 to 8.5 GHz
–85 to 10 dBm
–85 to 16 dBm
–85 to 12 dBm
–85 to 10 dBm
Max Leveled Power
30 to 300 kHz
300 kHz to 1 GHz
1 to 3 GHz
3 to 8.5 GHz
16 dBm
20 dBm
18 dBm
Slope from 18 dBm
(3 GHz) to 12 dBm
(8.5 GHz)
Specification (dB)
Description
30 to 300 kHz
300 kHz to 100 MHz 100 MHz to 2 GHz
2 to 4.5 GHz
4.5 to 8.5 GHz
Load match
14 (Typ.)
22
15
11
8
1. Frequency can be set down to 9 kHz
2. Source output performance on port 1 only. Port 2 output performance is typical.
3. Power can be set from –109 to 20 dBm