Corrected system performance – Atec Agilent-N5072A Series User Manual
Page 3
3
Corrected System Performance
The specifications in this section apply to measurements made with the Agilent E5072A network analyzer under the fol-
lowing conditions:
• No averaging applied to data
• Environmental temperature of 23 °C (± 5 °C) with less than 1 °C deviation from the calibration temperature
• Response and isolation calibration performed
Description
Specification
SPD
System dynamic range at test port
1, 2
(IF Bandwidth = 3 kHz)
30 to 300 kHz
65 dB (Typ.)
300 kHz to 10 MHz
82 dB
10 MHz to 6 GHz
98 dB
6 to 8.5 GHz
92 dB
(IF Bandwidth = 10 Hz)
30 to 300 kHz
90 dB (Typ.)
100 dB
300 kHz to 10 MHz
107 dB
115 dB
10 MHz to 6 GHz
123 dB
130 dB
6 to 8.5 GHz
117 dB
128 dB
Extended dynamic range at direct receiver access input
3
(IF Bandwidth = 10 Hz)
30 to 300 kHz
128 dB
300 kHz to 10 MHz
134 dB
10 MHz to 3 GHz
151 dB
3 GHz to 6 GHz
147 dB
6 to 8.5 GHz
145 dB
1. The test port dynamic range is calculated as the difference between the test port rms noise floor and the source maximum output power.
The effective dynamic range must take measurement uncertainty and interfering signals into account.
2. Typical performance might not be met from 60 to 70 kHz.
3. The direct receiver access input system dynamic range is calculated as the difference between the direct receiver access input noise fl oor and the
source maximum output power. The effective dynamic range must take measurement uncertainties and interfering signals into account.