Corrected system performance with calibration kit, Specification (db) – Atec Agilent-N5072A Series User Manual
Page 5
5
Figure 2. Transmission uncertainty (Specifi cation)
Figure 3. Refl ection uncertainty (Specifi cation)
Corrected system performance with calibration kit
Corrected system performance with type-N device connectors, 85032F calibration kit
Network analyzer : E5072A
Calibration kit
: 85032F (Type-N, 50 Ω)
Calibration :
full
2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (± 5 °C ) with < 1 °C deviation from
calibration temperature, isolation calibration performed.
Specification (dB)
Description
30 to 300 kHz (Typ.)
300 kHz to 10 MHz
10 MHz to 3 GHz
3 to 6 GHz
6 to 8.5 GHz
Directivity
49
49
46
40
38
Source match
41
41
40
36
35
Load match
49
49
46
39
37
Refl ection tracking
± 0.011
± 0.011
± 0.021
± 0.032
± 0.054
Transmission tracking
± 0.011
± 0.007
± 0.029
± 0.074
± 0.088
0.01
0.1
1
10
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
U
n
cer
tai
nt
y (
d
B
)
Transmission coefficient (dB)
Magnitude
30 KHz to 300 kHz
300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
0
0.01
0.02
0.03
0.04
0.05
0
0.2
0.4
0.6
0.8
1
U
n
cer
tai
nt
y (
li
n
ear
)
Reflection coefficient (linear)
Magnitude
30 KHz to 300 kHz
300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
0.1
1
10
100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
U
n
c
e
rt
a
in
ty
(
degrees)
Transmission coefficient (dB)
Phase
30 KHz to 300 kHz
300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
0
2
4
6
8
10
0
0.2
0.4
0.6
0.8
1
U
n
c
e
rt
a
in
ty
(
degrees)
Reflection coefficient (linear)
Phase
30 KHz to 300 kHz
300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz