beautypg.com

Front-panel jumpers – Atec Agilent-N5072A Series User Manual

Page 14

background image

14

Front-Panel Jumpers

Measurement Receiver Input (R1, R2, A, B)

Description

SPD

Typical

Noise Floor
30 kHz to 10 MHz
10 MHz to 8.5 GHz

–128 dBm/Hz
–145 dBm/Hz

(10Hz IFBW )
30 kHz to 10 MHz
10 MHz to 8.5 GHz

–118 dBm
–135 dBm

Maximum Input Level
(Rcvr R1/R2/A or B IN at 0.1dB typical compression)
30 to 300 kHz
300 kHz to 8.5 GHz

–15 dBm
–10 dBm

Damage Level

+15 dBm or ± 16 VDC

Coupler arm output

Description

SPD

Typical

Damage Level

+15 dBm or 0 VDC

Coupler through input

Description

SPD

Typical

Insertion Loss from Coupler
Thru to Test Port

Slope from 0.8 dB at
30 kHz to 3.5 dB at
8.5 GHz (0.32 x f [GHz] + 0.8) dB

Damage Level

+26 dBm or ± 35VDC

Source output

Description

SPD

Typical

Maximum Leveled Power
30 to 300 kHz
300 kHz to 1 GHz
1 to 3 GHz
3 to 8.5 GHz

+18 dBm
+20.5 dBm
+19 dBm
Slope from +19dBm at 3 GHz to
+16 dBm at 8.5 GHz

Maximum Input Level
(at Max Specifi ed Output Power)
30 kHz to 8.5 GHz

+20 dBm

Damage Level

+26 dBm or 0 VDC

Reference source output

Description

SPD

Typical

Damage Level

+15 dBm or 0 VDC