Stability

Description

Specification

SPD

Magnitude
30 to 300 kHz
300 kHz to 3 GHz
3 to 6 GHz
6 to 8.5 GHz

± 0.04 dB/°C
± 0.005 dB/°C
± 0.03 dB/°C
± 0.04 dB/°C

Phase
30 to 300 kHz
300 kHz to 3 GHz
3 to 6 GHz
6 to 8.5 GHz

± 0.8 deg/°C
± 0.1 deg/°C
± 0.4 deg/°C
± 0.8 deg/°C

Dynamic Accuracy

1

Description

Specification (dB)

Typical

Magnitude
10 dBm
–30 dBm
–100 dBm
–110 dBm

± 0.21 dB
± 0.045 dB
± 2 dB

± 3.0 dB

Phase
10 dBm
–30 dBm
–100 dBm

± 5 deg
± 0.3 deg
± 15 deg

1. Accuracy of the test port input power reading is relative to 10 dBm reference input power level.