Options, Calibration kits and ecal modules, Defi nitions – Atec Agilent-N5072A Series User Manual
Page 2: Boundary conditions
2
Options
This document provides technical specifi cations for the E5072A ENA network analyzer.
Calibration kits and ECal modules
This E5072A data sheet provides technical specifications for the following calibration kits and ECal modules.
For models not listed in this data sheet, please download the free Uncertainty Calculator from
www.agilent.com/find/na_calculator to generate the curves for your calibration kit.
E5072A-245
2-port with confi gurable test set, 30 kHz to 4.5 GHz
E5072A-285
2-port with confi gurable test set, 30 kHz to 8.5 GHz
85032F
Calibration kit
85033E
Calibration kit
85092C
Electronic calibration (ECal) module
85093C
Electronic calibration (ECal) module
Defi nitions
Specifi cation (spec.):
Warranted performance. All specifi cations apply at 23 ºC (± 5 ºC), unless otherwise stated, and 90 minutes after the
instrument has been turned on. Specifi cations include guard bands to account for the expected statistical performance
distribution, measurement uncertainties, and changes in performance due to environmental conditions.
Typical (typ.):
Describes performance that will be met by a minimum of 80% of all products. It is not guaranteed by the product warranty.
Supplemental performance data (SPD):
Supplemental performance data represents the value of a parameter that is most likely to occur; the expected mean or
average. It is not guaranteed by the product warranty.
General characteristics:
A general, descriptive term that does not imply a level of performance.
Boundary Conditions
In this data sheet, boundary conditions are given for the specifi cations. For example, system dynamic range is 98 dB with
the following boundary conditions.
Frequency: 10 MHz, IF bandwidth: 3 kHz
If the same boundary conditions fall under more than one category in a table, apply the best value.