Data transfer time, Cycle time vs. if bandwidth – Atec Agilent-N5072A Series User Manual
Page 22
22
Data transfer time
1
Number of Points
51
201
401
1601
SCPI over GPIB
64-bit fl oating point
5 ms
15 ms
28 ms
110 ms
32-bit fl oating point
3 ms
8 ms
15 ms
56 ms
ASCII
21 ms
78 ms
156 ms
618 ms
SCPI over 100 Mbps LAN (Socket)
2
REAL 64
1 ms
1 ms
2 ms
3 ms
REAL 32
1 ms
1 ms
1 ms
2 ms
ASCII
13 ms
47 ms
93 ms
363 ms
SCPI over 100 Mbps LAN (SICL-LAN)
REAL 64
3 ms
3 ms
4 ms
6 ms
REAL 32
3 ms
3 ms
4 ms
4 ms
ASCII
3 ms
7 ms
11 ms
37 ms
SCPI over USB
2
REAL 64
2 ms
2 ms
2 ms
3 ms
REAL 32
2 ms
2 ms
2 ms
3 ms
ASCII
3 ms
9 ms
17 ms
66 ms
SCPI over GPIB/USB (82357B)
REAL 64
8 ms
14 ms
23 ms
77 ms
REAL 32
7 ms
10 ms
14 ms
41 ms
ASCII
74 ms
285 ms
568 ms
2268 ms
COM
3
Variant Type
1 ms
1 ms
1 ms
1 ms
Cycle time vs. IF bandwidth
Condition: NOP=201, system error correction: OFF, Sweep mode: swep
IF BW
(Hz)
Cycle
Time
(ms)
IF BW
(Hz)
Cycle
Time
(ms)
IF BW
(Hz)
Cycle
Time
(ms)
IF BW
(Hz)
Cycle
Time
(ms)
IF BW
(Hz)
Cycle
Time
(ms)
10
19297
100
1931
1000
195
10000
20
100000
3
15
12865
150
1288
1500
130
15000
14
150000
3
20
9649
200
1091
2000
98
20000
11
200000
3
30
6433
300
645
3000
66
30000
8
300000
3
40
4825
400
484
4000
50
40000
6
400000
2
50
3861
500
388
5000
40
50000
5
500000
2
70
2735
700
276
7000
29
70000
4
1. Measured using a VEE Pro 7.0 program running on a 3.2 GHz Pentium 4 DELL Precision 370, Transferred complex S11 data,
using : CALC{1-36} : DATA : FDAT?.
2. USB Test and Measurement Class (TMC) interface that communicates over USB, complying with the IEEE 488.1 and IEEE 488.2 standards.
3. Measured using an E5072A VBA macro running inside the analyzer. Transferred complex S11 data