Analysis bandwidth

1

Maximum bandwidth

Option B25

Standard

25 MHz

10 MHz

Video bandwidth (VBW)

Range

1 Hz to 3 MHz (10 % steps, E24 series 24 per decade), 4, 5, 6, 8 MHz, and wide open

(labeled 50 MHz)

Accuracy

± 6 % (nominal)

Measurement speed

2

Standard

Local measurement and display update rate

4 ms (250/s) (nominal)

Remote measurement and LAN transfer rate

5 ms (200/s) (nominal)

Marker peak search

1.5 ms (nominal)

Center frequency tune and transfer (RF)

20 ms (nominal)

Center frequency tune and transfer (μW)

47 ms (nominal)

Measurement/mode switching

39 ms (nominal)

1. Analysis bandwidth is the instantaneous bandwidth available around a center frequency over which the input signal can be digitized for further

analysis or processing in the time, frequency, or modulation domain.
2. Sweep points = 101.