Indicated noise in cispr bw – Atec Agilent-N9038A User Manual
Page 12
12
Indicated noise in CISPR BW
Calculated from DANL data; EMI-AVG detector, 0 dB input attenuation; indicated RBW is CISPR RBW
RF Input 1; RF Input 2 to 1 GHz; RF Input 2 performance = RF Input 1 performance +11 dB
Typical including NFE
1
Preselector on, preamp off
20 Hz (1 Hz RBW)
3
100 Hz (10 Hz)
3
1 kHz (100 Hz)
3
9 kHz (200 Hz RBW)
100 kHz (200 Hz)
1 to 3 MHz (9 kHz)
3 to 30 MHz (9 kHz)
30 to 300 MHz (120 kHz)
300 to 600 MHz (120 kHz)
600 MHz to 1 GHz (120 kHz)
1 to 2 GHz (1 MHz)
2 to 2.5 GHz (1 MHz)
2.5 to 3 GHz (1 MHz)
3 to 3.6 GHz (1 MHz)
3.5 to 8.4 GHz (1 MHz)
8.3 to 13.6 GHz (1 MHz)
13.5 to 17.1 GHz (1 MHz)
17.0 to 20.0 GHz (1 MHz)
20.0 to 26.5 GHz (1 MHz)
+7 dBuV
2
+8 dBuV
2
+7 dBuV
2
–1.0 dBuV
–12.0 dBuV
–12.0 dBuV
–13.0 dBuV
–3.0 dBuV
–6.0 dBuV
–4.0 dBuV
+6.0 dBuV
+3.0 dBuV
+4.0 dBuV
+6.0 dBuV
+6.0 dBuV
+9.0 dBuV
+14.0 dBuV
+16.0 dBuV
+22.0 dBuV
Preselector on, preamp on
1 kHz (100 Hz RBW)
3
9 kHz (200 Hz RBW)
100 kHz (200 Hz)
1 to 2 MHz (9 kHz)
2 to 30 MHz (9 kHz)
30 to 600 MHz (120 kHz)
600 to 800 MHz (120 kHz)
800 MHz to 1 GHz (120 kHz)
1 to 2 GHz (1 MHz)
2 to 2.75 GHz (1 MHz)
2.75 to 3.6 GHz (1 MHz)
3.5 to 8.4 GHz (1 MHz)
8.3 to 13.6 GHz (1 MHz)
13.5 to 17.1 GHz (1 MHz)
17.0 to 20.0 GHz (1 MHz)
20.0 to 26.5 GHz (1 MHz)
–6 dBuV
2
–23.0 dBuV
–34.0 dBuV
–31.0 dBuV
–20.0 dBuV
–8.0 dBuV
–9.0 dBuV
–9.0 dBuV
+3.0 dBuV
–1.0 dBuV
+3.0 dBuV
–5.0 dBuV
–6.0 dBuV
–4.0 dBuV
+2.0 dBuV
+5.0 dBuV
1. Typical Indicated Noise including NFE = Typical DANL+ Bandwidth and Log corrrections-DANL improvement with NFE
2. No NFE factor at this frequency.
3. Specified for instruments with prefixes MY/SG5213 or greater. Nominal for instruments with earlier prefixes.